SAR REFERENCE DIPOLE CALIBRATION REPORT
Ref: ACR.262.4.20.MVGB.A
6 CALIBRATION MEASUREMENT RESULTS
6.1 RETURN LOSS AND IMPEDANCE IN HEAD LIQUID
Frequency (MHz)
Return Loss (dB)
Requirement (dB)
Impedance
750
-25.47
-20
55.2 + 1.5 j
6.2 RETURN LOSS AND IMPEDANCE IN BODY LIQUID
Frequency (MHz)
Return Loss (dB)
Requirement (dB)
Impedance
750
-27.25
-20
51.4 + 4.1 j
6.3 MECHANICAL DIMENSIONS
Frequency MHz
L mm
h mm
d mm
required
measured
required
measured
required
measured
750
176.0 ±1 %.
PASS
100.0 ±1 %.
PASS
6.35 ±1 %.
PASS
835
161.0 ±1 %.
89.8 ±1 %.
3.6 ±1 %.