SAR REFERENCE DIPOLE CALIBRATION REPORT
Ref: ACR.262.8.20.MVGB.A
2100
39.8 ±5 %
1.49 ±5 %
2300
39.5 ±5 %
1.67 ±5 %
2450
39.2 ±5 %
1.80 ±5 %
2600
39.0 ±5 %
1.96 ±5 %
3000
38.5 ±5 %
2.40 ±5 %
3500
37.9 ±5 %
2.91 ±5 %
7.2 SAR MEASUREMENT RESULT WITH HEAD LIQUID
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.
Software
OPENSAR V4
Phantom
SN 20/09 SAM71
Probe
SN 18/11 EPG122
Liquid
Head Liquid Values: eps’ : 41.1 sigma : 1.42
Distance between dipole center and liquid
10.0 mm
Area scan resolution
dx=8mm/dy=8mm
Zoon Scan Resolution
dx=8mm/dy=8m/dz=5mm
Frequency
1900 MHz
Input power
20 dBm
Liquid Temperature
21 °C
Lab Temperature
21 °C
Lab Humidity
45 %
Frequency
MHz
1 g SAR (W/kg/W)
10 g SAR (W/kg/W)
required
measured
required
measured
750
8.49
5.55
835
9.56
6.22
900
10.9
6.99
1450
29
16
1500
30.5
16.8
1640
34.2
18.4
1750
36.4
19.3
1800
38.4
20.1
1900
39.7
39.84 (3.98)
20.5
20.20 (2.02)
1950
40.5
20.9
2000
41.1
21.1
2100
43.6
21.9
2300
48.7
23.3
Page: 8/11
This document shall not be reproduced, except in full or in part, without the written approval of MVG. The information contained herein is to be used
only for the purpose for which it is submitted and is not to be released in whole or part without written approval of MVG.
Template_ACR.DDD.N.YY.MVGB.ISSUE_COMOSAR Dipole vH