Schematics

SAR REFERENCE DIPOLE CALIBRATION REPORT
Ref: ACR.262.11.20.MVGB.A
6 CALIBRATION MEASUREMENT RESULTS
6.1 RETURN LOSS AND IMPEDANCE IN HEAD LIQUID
Frequency (MHz)
Return Loss (dB)
Requirement (dB)
Impedance
2600
-34.32
-20
50.3 + 1.9 j
6.2 RETURN LOSS AND IMPEDANCE IN BODY LIQUID
Frequency (MHz)
Return Loss (dB)
Requirement (dB)
Impedance
2600
-24.15
-20
45.6 - 4.5 j
6.3 MECHANICAL DIMENSIONS
Frequency MHz
L mm
h mm
d mm
required
measured
required
measured
required
measured
750
176.0 ±1 %.
100.0 ±1 %.
6.35 ±1 %.
835
161.0 ±1 %.
89.8 ±1 %.
3.6 ±1 %.
Page: 6/11
This document shall not be reproduced, except in full or in part, without the written approval of MVG. The information contained herein is to be used
only for the purpose for which it is submitted and is not to be released in whole or part without written approval of MVG.
Template_ACR.DDD.N.YY.MVGB.ISSUE_COMOSAR Dipole vH