Datasheet

- IT / Multimedia Equipment
EN 62368-1
IEC 62368-1
UL 60950-1
UL 62368-1
- Medical Equipment
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents
www.tracopower.com/overview/tpp150a-j
Class I & II (Prepared): Reinforced Insulation
PD 2
OVC II
Safety Specifications
EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15 class B (internal filter)
FCC Part 18 class B (internal filter)
- Radiated Emissions
EN 55011 class A (internal filter)
EN 55032 class A (internal filter)
FCC Part 15 class A (internal filter)
FCC Part 18 class A (internal filter)
- Harmonic Current Emissions
EN 61000-3-2, class A
EN 61000-3-2, class D
- Voltage Fluctuations & Flicker
EN 61000-3-3
EN 55024 (IT Equipment)
EN 55035 (Multimedia)
EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge
Air:
EN 61000-4-2, ±15 kV, perf. criteria A
Contact:
EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field
EN 61000-4-3, 20 V/m, perf. criteria A
- EFT (Burst) / Surge
EN 61000-4-4, ±2 kV, perf. criteria A
L to L:
EN 61000-4-5, ±1 kV, perf. criteria A
L to PE:
EN 61000-4-5, ±2 kV, perf. criteria A
- Conducted RF Disturbances
EN 61000-4-6, 20 Vrms, perf. criteria A
- PF Magnetic Field
Continuous:
EN 61000-4-8, 10 A/m, perf. criteria A
- Voltage Dips & Interruptions
230 VAC / 50 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
60%, 5 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 250 periods, perf. criteria B
EMC Specifications
95% max. (non condensing)
- Operating Temperature
–25°C to +85°C
- Storage Temperature
–40°C to +85°C
- High Temperature
Depending on model
- Low Input Voltage
1.33 %/V below 100 VAC
See application note:
www.tracopower.com/overview/tpp150a-j
- Option 1
Forced air cooling (with external fan, 10 CFM)
- Option 2
Natural convection (20 LFM)
- Characteristic
Variable fan speed (temperature regulated)
- Output Voltage
12 VDC
- Output Current
500 mA max.
General Specifications
TPP 150A-J Series, 150 Watt
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com September 20, 2023