CALIBRATION
Calibration may be required when it is initially installed, if the scale is moved to a
substantial distance.
Calibration:
Single segment calibration (recommended):
1.
Turn on the scale and then turn it off.
2.
Press and hold [ZERO] and then turn on the scale, LCD will show SCALE on
first line, CAL-0 (flashed) on the second line and AD value on the third line.
3.
Press [SAMPLE/UNIT] to select calibration unit kg or lb.
4.
After the stable A/D value is displayed, press [ZERO] to calibrate ZERO, after
2 or 3 seconds, the second will show 0.
5.
Set calibration weight through input 0 to 9 and then place the calibration weight
you had set, Press [ZERO] to calibrate, after the stable indicator light, press
[ZERO] key, the second line of display will show 00000 and now calibration is
finished.
6.
Turn off the power, then turn on the power again, place some weight on the
platform to make sure whether weighing is correct, if not, repeat steps 1-6.
Linearity calibration (for scale technician or factory use):
1. Turn on the scale and then turn it off.
2. Press and hold [TARE] and then turn on the scale, LCD will show LINE
on first row,CAL-0 (flashed) on the second line and AD value on the third line.
3. When the AD value is stable, press [ZERO] to calibrate ZERO, after 2 or 3
seconds, the second line will show 1.0000.
4. Place 1kg weight on the platform , and press [ZERO] when the stable A/D value
is displayed, after 2 or 3 seconds, the display will show 2.0000.
5. Place 2kg weight on the platform , and press [ZERO] when the stable A/D value
is displayed, after 2 or 3 seconds, the display will show3.0000.
6. Place 3kg weight on the platform , and press [ZERO] when the stable A/D value
is displayed, after 2 or 3 seconds, the display will show0.0000 and now calibration is
finished.
7. Turn off the power, then turn on the power again, place some weight on the
platform to make sure whether weighing is correct, if not, repeat steps 1-6.