Calibration & Connectivity TS-700 Series Digital Indicators This document supplements the User’s Guide TS-700 MS TS-700 SS TS-700 WB Full Function/Advanced Function Digital Indicator www.trinerscale.com Version 4.
Triner Scale & Mfg.Co., Inc. 8411 Hacks Cross Road Olive Branch, MS 38654 (800) 238-0152 www.trinerscale.com © 2013 Triner Scale. All rights reserved.
Calibration & Connectivity | TS-700 Series Table of Contents Specifications...................................................2 Load Cell Connections .....................................3 Serial Port Connections ...................................4 Calibrating a TS-700 Series Indicator...............6 Data Output Formats .....................................12 Error Codes ...................................................15 Factory Default Settings.................................
Calibration & Connectivity | TS-700 Series Specifications TECHNICAL PARAMETERS Resolution ...................... Display: 30,000 ADC ................................ 2,000,000 Zero stability error ........... TK0 < 0.1μV//K Span stability error .......... TKspn < ± 6 ppm//K Sensitivity (internal)......... 0. 3 μV /d Input voltage ................... -30~30mV DC Excitation circuit .............. 5 VDC Load Cell Max................. 8, 350Ω AC power ........................ AC100~250V Operation temp ........
Calibration & Connectivity | TS-700 Series Load Cell Connections The indicator can connect with a maximum of eight (8) 350Ω load cells, 4 wire or 6 wire configurations. Excitation voltage for the load cell is 5VDC, the maximum output current is 120mA.
Calibration & Connectivity | TS-700 Series Serial Port Connection RS232 CONNECTION TS-700 MS The RS232 DB9 serial port can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC. NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Calibration & Connectivity | TS-700 Series TS-700 SS The 3-wire terminal block located on the main board can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC. NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Calibration & Connectivity | TS-700 Series Calibrating a TS-700 Series Indicator NOTICE: The TS-700 SS/MS/WB are NTEP Class III Legal for Trade approved and certified. Unlocking the calibration parameters is not allowed by persons other than qualified scale technicians. DEFAULT SETTINGS CAUTION Placing the indicator in calibration mode enables access to the Default Settings Parameter, CO7.
Calibration & Connectivity | TS-700 Series Calibration Switch, TS-700 MS: Remove the screws and the back plate of the enclosure. A white-capped calibration switch is located at the bottom left corner of the circuit board. Press the switch fully in and release. This will place the switch in the up position, allowing access to the calibration parameters. Loosely reattach the rear plate. Calibration Switch, TS-700 SS: Remove the screws and the back plate of the enclosure.
Calibration & Connectivity | TS-700 Series calibration parameters, proceed as follows: CALIBRATION PROCEDURE 1) Power on the indicator. 2) Press the PRINT and HOLD keys. “C” will display. Release the keys. 3) “C01” will display, with the 1 flashing. The indicator is now is Parameter Settings mode. Navigating in Settings Mode When the indicator is in settings mode, refer to the graphic printed below the key to determine the keypad key’s function.
Calibration & Connectivity | TS-700 Series PARAMETER FUNCTION OPTIONS C01 Default weighing unit 1 = kg 2 = lb Description: With C01 flashing, press the key, then use the up or down arrow keys to change the parameter setting as needed. key to accept Press the the setting and continue on to the next parameter.
Calibration & Connectivity | TS-700 Series C03 Graduations 1 = one unit 2 = two units 5 = five units 10 = ten units 20 = twenty units 50 = fifty units Description: With C03 flashing, press the key, then use the up or down arrow keys to change the parameter setting as needed. key to accept Press the the setting and continue on to the next parameter.
Calibration & Connectivity | TS-700 Series C06 Calibration 0 = skip calibration 1 = proceed with calibration Description: With C06 flashing, press the key, then use the up arrow key to change the parameter setting to 1. key to Press the proceed. “Span” will briefly display, then all digits will display. Load the weighing platform, then adjust the displayed digits to match the weight on the platform, i.e, 4,000. key to Press the proceed. The indicator will calibrate, then display “CalEnd”.
Calibration & Connectivity | TS-700 Series Data Output Formats REMOTE DISPLAY FORMAT State A Bits: 0,1,2 0 1 2 1 0 0 0 1 0 1 1 1 0 0 Bits: 3,4 0 0 1 1 0 1 1 0 Decimal point position XXXXXX0 XXXXXXX XXXXX-X XXXX-XX XXX-XXX Division X1 X2 State B BITS FUNCTION Bits0 Gross = 0, net=1 Bits1 Symbol: positive = 0,negative =1 Bits2 Overload (or under zero)=1 Bits3 Dynamic = 1 Bits4 Unit: lb=0, kg=1 Bits5 Bits6 Constant 1 Constant 0 12
Calibration & Connectivity | TS-700 Series State C Bit1 0 0 1 Bit 3 Bit2 0 0 0 Bit0 0 1 0 Bit 4 Bit 5 Bit 6 unit Kg or lb g t printing=1 Extend display=1 Constant 1 Constant 0 PC COMPUTER CONTINUOUS SENDING FORMAT , S1 CR , S2 S3 Data S1: Weight status: ST = standstill US = not standstill OL = overload S2: Weight mode: GS = gross mode NT = net mode S3: Weight of positive and negative, “+” or ” –“ S4: “kg” or “lb” Data: Weight value, including decimal point CR: Carriage return LF: Line
Calibration & Connectivity | TS-700 Series SERIAL INTERFACE RECEPTION COMMAND RS232COM serial interface can receive simple ASCII command. Command word and role as follows: Command T Z P R Name TARE ZERO PRINT G.W/N.W Function Save and clear tare Zero gross weight Print the weight Read gross weight or net weight C Kg/lb Kg/lb conversion G G.W Check gross weight at net weight mode Data Format Print Format ID.NO...................004 (Serial No.) Date: ....................XX.XX. XX (yy.mm.dd) Time:.
Calibration & Connectivity | TS-700 Series Error Codes ERROR CAUSE uuuuuu nnnnnn SOLUTION 1. Overloaded condition. 2. Wrong connection at load cell. 3. Load cell has quality problem. 1. Reduce the weight. 2. Check load cell connection 3. Inspect load cell. Check the input and output. 1. Calibration is incorrect. 2. Bad connections. 3. Load cell has quality problem. 1. Check scale is resisted or not, foot is kept level or not. 2. Check load cell connection. 3.
Calibration & Connectivity | TS-700 Series Factory Default Settings PARAMETER INSTRUCTION DEFAULT C01 Calibration 1 C02 Decimal digits 0 C03 Resolution 1 C04 Max.
Calibration & Connectivity | TS-700 Series C23 Zero tracking time C24 C25 C26 C27 C28 C29 C30 C31 C32 C33 C34 C35 C36 C37 C38 C39 Overload range Negative range Standstill time Standstill range Dynamic filter Noise filter Print format Analog signal options 4~20mA testing Relay output setting Reserved Reserved Calibration location gravity Destination gravity Version No. check Reserved menu 17 1 9 10 1 2 0 2 0 1 4 1 0 6 9.7936 9.