Setup Guide

10
ENGLISH
Dispose of the CaviWipe.
Step 4: Disinfect
Take a CaviWipe.
Wet the surface of the TRIOS scanner body and
mirror housing with the CaviWipe for a minimum
of 120 seconds to disinfect.
Allow to dry.
Dispose of the CaviWipe.
Place TRIOS protection tip on TRIOS scanner.
CAUTION
The TRIOS scanners window is a delicate optical component. Its clean and
undamaged surface is critical for the scan quality. The window should be
handled with care. Make sure it is free from lint, stains, and other kinds of dirt.
CAUTION
Never submerge the TRIOS scanner into any cleaning or disinfection solution.
CAUTION
Before returning the TRIOS scanner to its holder, ensure both are dry.
IMPORTANT NOTICE
The TRIOS scanner shell might get yellow-toned over a long period of use.
7.2 Pod: Cleaning and Disinfection
It is recommended to disassemble the pod in order to thoroughly clean it.
Follow the instructions below (step 1-6)
Step 1: Disassemble pod
Wearing gloves, remove the rubber insert from
the pod (holder) by pressing down on the middle
using a thumb and gripping the outer edge with
the index finger.
Step 2: Remove gross contaminants from rubber insert
Take a CaviWipe.
Wipe the surface of the top rubber insert to remove
any gross contaminants.
Dispose of the CaviWipe.
Step 3: Remove gross contaminants from metal holder
Take a CaviWipe.
Wipe the surface of the metal holder to remove any
gross contaminants.
Dispose of the CaviWipe.
Step 4: Disinfect rubber insert
Take another CaviWipe.
Wet the surface of the rubber insert for minimum
120 seconds to disinfect.
Dispose of the CaviWipe.
Step 5: Disinfect metal holder
Take a CaviWipe.
Wet the surface of the metal part for minimum
120 seconds to disinfect.
Dispose of the CaviWipe.
Step 6: Assemble pod
Allow both parts to dry. Assemble the pod (holder).
7.3 TRIOS Scanner Tip: Cleaning and Sterilization
TheTRIOS scanner tip isthepart of theTRIOS scanner that isinserted intothepatient’s
mouth during scanning. TheTRIOS scanner tip canbe re-used up to 60times. The TRIOS
scanner tip should be cleaned and sterilized prior to its first use on patients and after each
use on a patient toavoid cross-contamination.
Step 1: Clean TRIOS scanner tip
Wearing gloves, cleantheTRIOS scanner tip
manually using detergent diluted according
to manufacturer’s instructions.
Apply the detergent directly on a soft bristled brush.
Wash the TRIOS scanner tip under running lukewarm
(30-40 °C) utility water.
Inspect thewindowof theTRIOS scanner tip
after cleaning. Ifthewindowappears withstains,
smudged orwithamilky haze, repeat this step.
Step 2: Rinse and dry TRIOS scanner tip
Rinsewithlukewarm (30-40 °C) tap water.