Setup Guide

5
ENGLISH
4.11 Starting and Shutting Down the TRIOS System
Please follow the steps below to start TRIOS system for the first time:
Step 1: Turn the PC on.
Start 3Shape software.
Step 2: Log on using a default TRIOS Operator.
Step 3: Configure and use your TRIOS system.
Follow on-screen instructions forconnecting tothewireless TRIOS scanner.
When using theTRIOS system, theuser is guided bymessages shown intheTRIOS software.
The TRIOS scanner has an LED ring which indicates the status of the scanner:
A rotating blue light indicates that the TRIOS scanner
is waiting for the TRIOS software to connect to it.
Three blue blinks indicate that the TRIOS scanner is
searching for a scanner application to connect to.
A constant blue light indicates that the TRIOS
scanner is connected to the system, but no
scanner tip is mounted.
A rotating green light indicates that the TRIOS
scanner is heating the TRIOS scanner tip.
A constant green light indicates that the TRIOS
scanner is ready to scan.
A rotating white light indicates that the TRIOS
scanner is waiting for scan data.
A constant white light indicates that the TRIOS
scanner is scanning. The light intensity decreases
if the TRIOS scanner loses registration.
Solid yellow indicates that the battery is low.
Scanner operation
The scanner has two buttons for operation; the button closests to the tip controls scanning
operation, the button farthest from the tip controls navigation.
Power saving
The TRIOS scanner has integrated power control, and will go into sleep mode when not used,
unless the software is on the Scan page.
4.12 Starting TRIOS Online Help
The TRIOS online help can be started from the TRIOS software by clicking on the Help
button and choosing User Manual.
4.13 Storage, Operating and Transport Conditions
Operating Conditions
Ambient temperature range: +15 – +26 °C (+59 – +78.8 °F)
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
The TRIOS scanner should be used in an environment maintaining a
consistent room temperature.
Storage Conditions
Ambient temperature range: -10 – +50 °C (+14– +122 °F)
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
Transport Conditions
Ambient temperature range: -10 – +50 °C (+14– +122 °F)
Relative humidity: 10 – 85 % (Non-condensing)
Atmospheric pressure: 620 – 1200 hPa
4.14 Technical Specifications
Light source
specifications
Donot view directly withoptical instruments, such asspectacle
microscopes. Theproduct complies withEN (IEC) 62471
(Photobiological safety oflamps andlamp systems andUL-standard
US code offederal regulations, 21CF 1040.10 and1040.11.
Sterilization Sterilize by autoclave (see section 7 for autoclave specifications)
Power input C2BP: Battery charger
Power input: 100-240 VAC 0.6 A, 50/60 Hz
Protection class: IP20
Insulation class Class II
L1-2 scanner. Wireless, battery powered
3Shape TB-04 Battery:
3,6 V, 3300 mAh, 12 Wh
Wireless TheTRIOS scanner contains an IEEE 802.11ac module
Wireless
Transmit Power
Maximum 100 mW
Wireless
Frequency Band
2.401-2.483 GHz
5.15-5.725 GHz (the actual frequencies are dependent on local
regulations and the configuration of the product.)
IT Network equipment
Network
characteristics
802.11ac (5GHz)
Capacity for at least 40 Mbps per scanner.
Network
configuration
Bi-directional traffic permitted between TRIOS scanner and PC.
mDNS enabled (required for scanner discovery).
Redirection of traffic must be disabled.
Channel width set to at least VHT40 (40MHz).
Ports used:
TCP ports (Scanner): 20-21, 80 and 23795
UDP Ports (Scanner): 58218 and 58220-58230
UDP Ports (PC and Router): 67 (DHCP), 5353 (mDNS)
Security
specification
WPA2-PSK authentication
IMPORTANT NOTICE
If you connect your TRIOS intraoral scanner system to your IT Network
equipment, it is strongly recommended to apply network security
measures and to analyse, evaluate, and control the IT network.
4.15 Electromagnetic Compability
The TRIOS system has been tested according Medical electrical equipment – Part 1-2:
General requirements for basic safety and essential performance– Collateral Standard:
Electromagnetic disturbances– Requirements and tests, EN/(IEC) 60601-1-2 ed. 4.0, 2015.
WARNING
Medical electrical equipment needs special precautions regarding EMC
and needs to be installed and put into service according tothe EMC
information provided in this document.
IMPORTANT NOTICE
Portable and mobile RF communications equipment can affect medical
electrical equipment.