5.3.7 Disconnect the test leads from the circuit under test and
return the Range Switch to the OFF position.
5.4 Measuring Resistance
5.4.1 Voltage Sensing Mode
Before measuring resistance, the circuit to be tested must be
checked for a voltage which is present as this may present a safety
hazard and/or give a false reading. This is facilitated through the
Ohmmeter's voltage sensing mode.
5.4.1.1 Refer to Section 3.0, Preparation For Use, and Section 4.0,
General Operation, before attempting any measurements.
5.4.1.2 Select the desired Ohms range by rotating the Range Switch
to the desired position. The position of the Code Function Switch
and the Signal Select Switch does not matter.
5.4.1.3 Attach the test leads to the Test Set. The red lead is con-
nected to the VOM Jack, and the black lead is connected to the
COM Jack.
5.4.1.4 Zero the Ohmmeter by shorting the test leads together and
depressing and holding the Push For Ohms Switch. The Pointer will
deflect. Use the Ohms Adjust Control to set the Pointer on the Ohm-
meter Zero Mark. If this is not possible, then the D-cell battery needs
replacement. Refer to Section 3.7. When zeroed, release the Push
For Ohms Switch.