Use Instructions

22
Lancing Device Cleaning and Disinfecting
Clean the lancing device immediately after getting any blood on it or if it is dirty.
Remove lancet from lancing device before cleaning.
Clean and disinfect the lancing device at least once a week. The lancing device may be
cleaned and disinfected once a day for a 3-year period.
Use only PDI Super Sani-Cloth® Wipes (EPA Reg. No. 9480-4). Other disinfectants have
not been tested and may damage the lancing device.
PDI Super Sani-Cloth Wipes have been tested on the lancing device for a total of 1,095
cleaning and disinfecting cycles, which is equal to cleaning and disinfecting the lancing
device once a day for a 3-year period.
Important! Stop using the lancing device and contact Customer Care for assistance, if:
markings on lancing device are coming off
trigger button hard to push
end cap does not go back on
arming barrel does not click when gently pulled back
To Clean the Lancing Device:
1. Wash hands thoroughly with soap and water.
2. Use fresh PDI Super Sani-Cloth Wipes to wipe the entire outside of the lancing device
using three circular wiping motions with moderate pressure. Discard used wipes.
To Disinfect the Lancing Device:
1. Always clean the lancing device immediately before disinfecting (see
instructions above).
2. Use fresh PDI Super Sani-Cloth Wipes to wipe the lancing device. Make sure that all
outside surfaces of the lancing device remain wet for 2 minutes.
3. Let the lancing device air dry thoroughly before using.
4. Replace the end cap, gently pull back the arming barrel and press the trigger button.
You will hear a click if the lancing device is functioning properly.
5. Wash hands thoroughly when finished.
Call Customer Care at 1-800-803-6025, Mon.–Fri., 8AM–8PM EST