Lancing Device Cleaning and Disinfecting
• Clean the lancing device immediately after getting any blood on it or if it is dirty.
• Remove lancet from lancing device before cleaning.
• Clean and disinfect the lancing device at least once a week. The lancing device may be
cleaned and disinfected once a day for a 3-year period.
• Use only PDI Super Sani-Cloth® Wipes (EPA Reg. No. 9480-4). Other disinfectants have
not been tested and may damage the lancing device.
• PDI Super Sani-Cloth Wipes have been tested on the lancing device for a total of 1,095
cleaning and disinfecting cycles, which is equal to cleaning and disinfecting the lancing
device once a day for a 3-year period.
Important! Stop using the lancing device and contact Customer Care for assistance, if:
• markings on lancing device are coming off
• trigger button hard to push
• end cap does not go back on
• arming barrel does not click when gently pulled back
To Clean the Lancing Device:
1. Wash hands thoroughly with soap and water.
2. Use fresh PDI Super Sani-Cloth Wipes to wipe the entire outside of the lancing device
using three circular wiping motions with moderate pressure. Discard used wipes.
To Disinfect the Lancing Device:
1. Always clean the lancing device immediately before disinfecting (see
instructions above).
2. Use fresh PDI Super Sani-Cloth Wipes to wipe the lancing device. Make sure that all
outside surfaces of the lancing device remain wet for 2 minutes.
3. Let the lancing device air dry thoroughly before using.
4. Replace the end cap, gently pull back the arming barrel and press the trigger button.
You will hear a click if the lancing device is functioning properly.
5. Wash hands thoroughly when finished.
Call Customer Care at 1-800-803-6025, Mon.–Fri., 8AM–8PM EST