Use Instructions

28
System Specications
Result Range: 20–600 mg/dL
Sample Size: Minimum 0.5 microliter (0.5 μL)
Sample:
Fresh capillary whole blood, venous whole blood collected in
an EDTA
blood collection tube, or control solution
Test Time: About 11 seconds
Result Value: Plasma values
Assay Method: Amperometric
Power Supply: One CR2032 3V lithium battery (non-rechargeable)
Battery Life: Approximately 1000 tests or 1 year
Automatic shut-off: After two minutes of non-use
Weight: 1.7 oz.
Size: 3.44” x 2.16” x 0.69”
Memory Size: 1000 blood glucose and control test results with date and time
Operating Range (Meter & Test Strips) (Use within specified environmental conditions only):
Temperature: 50°F–104°F
Humidity: 10%–90% (non-condensing)
Hematocrit: 20%–60%
Altitude: 10,150 feet
Chemical Composition
Test Buddy Test Strips:
Glucose dehydrogenase-FAD (Aspergillus species), mediators, buffers and stabilizers.
Test Buddy Control Solution:
Water, d-glucose, buffers, viscosity enhancing agent, salts, dye, and preservatives.
PetMeterBook.indd 28PetMeterBook.indd 28 11/25/19 16:0211/25/19 16:02