Product Manual

LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’
is displayed when the temperature being measured is outside of the settings of HAL and
LAL.
‘Er2’ is displayed when the thermometer is exposed to rap
is displayed when the ambient temperature exceeds 0
o
C (32°F) or +50
o
C (122°F).
The thermometer
should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.
Error 5~9, for
all other error messages it is necessary to reset the thermometer. To reset it, turn the
instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn
on. If the error message remains please contact the Service Department for further assistance.
」「
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are not
possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries. Please note:
It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item Non-contact Infrared Scan function
Thermocouple Probe Scan function
(K type; probe not included.)
Measurement Range
-60 to +625
o
C (-76 to +1157
o
F)
-64 to +1400°C (-83.2 to +1999°F)
Operating Range 0 to +50°C (32 to +122°F)
Accuracy
(Tobj=15~35°C, Tamb=25°C)
±1.0°C (1.8°F)
+/-1% of reading or 1°C (1.8°F) whichever
is
greater (Test under Tamb=23±6°C)
Accuracy(Tamb=23±3°C)
Tobj=0~625°C:+/-
2% of reading or 2°C (4°F) whichever is
greater, Tobj= -60~0
o
C: +/-(2°C+0.05/degree)
Emissivity Range 0.95 default – adjustable 0.1 to 1 step .01
Resolution (-9.9~199.9°C) 0.1°C/0.1°F
Response Time (90%) 1sec
Distance:Spot 16:1(90% energy covered)
Battery Life Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions 46.0 x 143.0 x 184.8mm(1.81×5.63×7.28 inch)
Weight 240 grams(8.5 oz) including batteries (AAA*2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10
o
C (18
o
F).