5.6 Probe Tip Replacement
The probe tip is replaceable. It can be unscrewed from the shaft
by using pliers to grip the shaft and cone respectively. The cone
is attached with removable type thread locker, so some force
may be required to loosen it. The probe tip should be replaced
when the diameter of the cone’s base measures less than
0.490in.(12.45mm) diameter.
6.0 Troubleshooting & Maintenance
Errors: If an error message (Err2 or Err3) is displayed during the
zero process, then retry the zero button and assure the meter’s
shaft and tip are hanging vertically below the meter and not
touching anything. If the error message continues to show after
repeated zeroing attempts, then the meter must be serviced.
Cleaning: Clean meter head and probe shaft with a damp cloth
when required. Store unit clean.