Product Overview

WWW.UEITEST.COM USA 1.800.547.5740
CANADA 1.877.475.0648
Made in CHINA
AC 400A TRMS Clamp Meter
DL469
Includes
Test leads, w/alligator clips
Zippered pouch
Batteries 2 (AAA)
Manual
Specifications
Downloads
Manual
Data Sheet
AC Amps - Jaw Input
Range Resolution Accuracy Overload Protection
40A 0.01A ±(3.0% +10 dgts)
600V RMS
400A 0.1A ±(2.5% +10 dgts)
True RMS Frequency range: Sine 50Hz to 400Hz
Frequency width: 60Hz to 400Hz: 5% to 95%, Frequency width: 400Hz to 4kHz: 15% to 85%
AC Volts (45Hz to 400Hz)
Range Resolution Accuracy Overload Protection
400mV 0.1mV
±(1.0% +8 dgts)
1000V RMS
4V 1mV
40V 10mV
400V 100mV
750V 1V ±(1.2% +8 dgts)
True RMS Frequency range: Sine 50Hz to 400Hz, Square 50Hz to 170Hz, Sine 50Hz to 400Hz
Bandwidth: Sine = 0.5% error at 1.5kHz (max)
Bandwidth: Square = 0.5% error at 0.1kHz (max)
Bandwidth: Triangle = 0.5% error at 1.2kHz (max)
DC Volts
Range Resolution Accuracy Overload Protection
400mV 0.1mV
±(0.8% +5 dgts)
1000V RMS
4V 1mV
40V 10mV
400V 100mV
600V 1V ±(1.0% +5 dgts)
DC Low Amps
Range Resolution Accuracy Overload Protection
400µA 0.1µA
±(1.2% +3 dgts) 2000µA/600V RMS
2000µA 1µA
Resistance
Range Resolution Accuracy Overload Protection
400Ω 0.1Ω
±(1.0% +5 dgts)
600V RMS
4kΩ
40kΩ 10Ω
400kΩ 100Ω
4MΩ 0.001MΩ
40MΩ 0.01MΩ ±(1.5% +5 dgts)
Diode Test
Range Open Circuit Test Current Overload Protection
4.0V <3.0V DC 1.30mA 600V RMS
Continuity
Open Circuit Response Time Overload Protection
<1.0V <50ms 600V RMS
Non-Contact Voltage
On Voltage
Approx. 25V AC