METER PROBLEM SOLVING
  If any problems are not solved with these solutions, contact us or an authorized repair   
 center.
Fault symptom  Causes / Solutions
CO or CO2 too low  • Air leaking into probe, tubing, water trap
  or connectors.
CO2 reading (- - - -)  • Meter was stored in a cold environment and 
   is not at normal working temperature.
Batteries not holding charge  • Batteries exhausted.
Meter does not respond to flue gas  • Particle filter blocked.
  • Probe or tubing blocked.
  • Pump not working or damaged with 
   contaminants.
Net temperature or Efficiency  • Ambient temperature set wrong during                             
calculation incorrect.  Automatic Calibration.
Flue temperature readings erratic  • Temperature plug reversed in socket.
  • Faulty connection or break in cable or plug.
T flue displays (- - - -)  • Probe not connected or faulty
X-Air, EFF display  • CO2 reading is too low.
 (- - - -) 