Service manual
4-1
Contents
Maintenance
Overview
Cleaning/Disinfecting
CAUTION:
• Observe precautions for handling electrostatic-sensitive devices!
!
• Never touch electrostatic-sensitive electronic components without following proper
anti-static procedures, including the use of an ESD wrist band and mat. An
electrostatic discharge from your fingers can permanently damage electronic
components.
• All static-sensitive electronic components are packaged in static-shielding bags.
Retain the bag for repackaging the component, should you need to store it or
return it to Spacelabs Medical for any reason.
• Cleaning, preventive maintenance, and safety checks should be performed
annually and following any product disassembly/assembly. Preventive
maintenance and safety checks must be performed by trained personnel only.
WARNING:
• Use only recommended cleaning solutions, or you may void the
manufacturer’s warranty.
• Harsh chemical agents degrade plastics and will compromise the safety of
the device.
• Disconnect the equipment from the patient and the electrical supply before
cleaning.
• Do not allow liquid to enter the interior of the module or monitoring
equipment.
• Do not immerse the equipment or cables in water or cleaning solutions.
• Do not autoclave.
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Preventive Maintenance (PM) Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
90363 Disassembly Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
90364 Disassembly Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20