Service manual

Ultraview 1500/1600 Monitors — Service Manual
4-10
Filter and Fan (90363 Only)
To clean the fan and filter:
1. Remove the filter cover from the rear of the Ultraview 1500 monitor using a small, flat head
screwdriver to gently pry off the cover.
2. Clean the filter by washing it with water and mild soap (or replace it).
3. Allow the filter to dry.
4. Reinstall the filter and the fan filter cover.
5. Verify that the fan is working properly and that it is drawing air into the display assembly.
Touchscreen (90363 Only)
The touchscreen does not have a separate shield over its face. Clean the screen with a soft cloth
moistened with either 70% alcohol or with soapy water.
90363 Disassembly Procedures
Tools Required
Static mat
Static wrist strap
#1 and #2 Phillips screwdrivers (#2 driver > 7.5 in. [19 cm] shaft)
13/16 in. (4.75 mm) nut driver
Standard flat screwdriver
Top cover removal tool (P/N 003-0085-00)
!
Depending upon the hospital environment, an air filter may need to be cleaned
more than once per year. It is recommended that, in the beginning, hospitals
inspect the air filter at monthly intervals until an appropriate cleaning cycle is
established.
!
Do not allow liquid to enter the monitor.
CAUTION:
Before disassembly, switch the main power OFF and disconnect the AC
cord from the wall outlet.