Specifications
WA-1100/WA-1600 Wavemeter Operating Manual 3 Remote Communications
08115-M-00 Rev E 3-21
10 1024 Fringe counter overflow.
11 2048 Fringe counter invalid.
12 4096 The input laser fringe strength is too high.
13 8192 The reference laser % of peak power is out of
range. (WA-1600 only)
SYSTem Subsystem
The SYSTem subsystem supplies information and status about the SCPI command parser
and allows control over the instrument's front panel.
SYSTem Subsystem Functions
:ERRor?
Description Reads error strings from the SCPI Error Queue. If the Error Queue has
any entries, the Error Queue bit is set in the Status Byte. The
instrument has a 30 entry, first-in, first-out queue. Repeatedly sending
the query :SYST:ERR? returns the error numbers and descriptions in
the order in which they occurred until the queue is empty. Any further
queries return 0, "No error" until another error occurs. A table of
possible errors is shown below.
Query Response <integer>, <string> (ex. –104, “Data type error”)
Examples :SYST:ERR?
SCPI Errors
Error Number Description
0 No error
-101 Invalid character
-102 Syntax error
-103 Invalid separator
-104 Data type error
-220 Parameter error
-221 Settings conflict
-222 Data out of range
-230 Data corrupt or stale