To program the Flow Oracle 160 for Example 1 V-Notch Weir by using
the Quick Setup menu proceed as follows.
If required access the Program Mode
Key in the passcode 1997 and press ENTER
Using the ‘right’ arrow key go to the Quick Setup menu press ENTER and
as prompted, by the questions, select the relevant option and press ENTER.
Question
Option
PMD Type
1 = Exponent
Exponent
6 = V notch.
Calculation
2 = Ratiometric.
No. of Alarms
0 = No Alarms
Xducer
1 = dB Mach3
Volume Units
6
=
Mil. USG
Time Units
4
= Per
Day
Measnt. Units
5
=
inches
Empty Level
40 inches
Minimum Head
0.000 m
etres
Maximum Head
15 inches
Total Enable
1 = On
Totaliser ( R )
1
=
Yes
Total Multiplier
1
=
/
1000
Maximum Flow
2.2
Programming is now complete and the unit can now be returned to the run
mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Flow Oracle 160 will return to Run Mode.