Instruction manual

Page 112
Universal Calculations
Universal CalculationsUniversal Calculations
Universal Calculations
Point of Measurement
The transducer must be above the maximum head P704 by at least the near
blanking distance P107.
For all Universal calculation applications the point at which the head is
measured should be chosen such that the surface of the liquid is not effected
by turbulence.
Calculations
Universal Head Vs Flow
A
BSOLUTE
If the flow calculation is to be absolute P702 = 1 the flow will be calculated
using the formula: q = q(h)
Where: q = flowrate
q(h) = flowrate for head
The desired number of Breakpoints, (P730 - P793) are to be entered in
pairs in values of head and corresponding flow. (Minimum of 2 pairs of
Breakpoints is required).