Instruction manual
Operation
Features of the menu screen
2-70 WX Ultra Series Thermo Scientific
2. Deleting the centrifuge reservation
You can cancel the operation schedule that is already reserved. This cancellation of reservation is
limited to the reserving person proper (with the same ID Code).
The reserved operation schedule will be deleted from the centrifuge reservation list after the lapse of the
scheduled date/time regardless of operation/non-operation of the centrifuge.
3 Enter an ID Code that is already regis-
tered. (4-digit number)
4 Select “Reserve”.
5 Use cursor keys to move the cursor to
the position at which you want to
change the date/time of reservation,
and enter a number by using numeric
keys.
Note Enter a time later than the current time.
6 At the desired date/time of reserva-
tion, press the ENTER key.
ID code
•The User Name
corresponding to the ID
Code is displayed. (In this
example, ABC lab. is
displayed.)
• The current date/time is
displayed at the reservation
time of the specified
reservation No.3.
Step Key operation Screen display and considerations
1 On the Menu Screen, select a centri-
fuge operation schedule.
• The Menu Screen switches
to the Centrifuge
Reservation screen.