Operating instructions

Loading mechanism of the weighing cell
13
Send K 1 (K 1<CRLF>)
Answer: e.g. K A<CRLF>
Send DW (DW<CRLF>)
Answer: e.g. DW A<CRLF>
Send D "end & print" (D "end & print"<CRLF>)
Answer: e.g. D A<CRLF>
Event: e.g. K 7<CRLF>
Print key
Press
print key
Send WS (WS<CRLF>)
Answer: e.g. WS A 0<CRLF>
If answer=WS 0 then go forward, else send WS again
Create SecondWeight (weight on pan)
Create and display results
If last probe then go forward, else go back to loop start
Send SU (SU<CRLF>)
Answer: e.g. 10.12 mg<CRLF>
If answer=WS A 2 then go forward, else send WS again
Send WS (WS<CRLF>)
Answer: e.g. WS A 2<CRLF>
Send "WS 1" (WS 1<CRLF>)
Answer: e.g. WS A<CRLF>
Send D "unload & print" (D "unlo. & print"<CRLF>)
Answer: e.g. D A<CRLF>
Send K 3 (K 3<CRLF>)
Answer: e.g. K A (K A<CRLF>)
remove weight
and press
print key
Print key
Event: e.g. K 7<CRLF>
Send "WS 0" (WS 0<CRLF>)
Answer: e.g. WS A<CRLF>
If answer=WS A 1 then go forward, else send WS again
Send WS (WS<CRLF>)
Answer: e.g. WS A 1<CRLF>
{unlo. = unload}
PC / Software UMX/MX Automated
Open comport
Send SI (SI<CRLF>)
Answer (e.g. "0.00003 g"<CRLF>)
Flush answer (only for communication check)
If communication ready then go forward, else stop program
Send "WS 1" (WS 1<CRLF>)
Answer: e.g. WS A<CRLF>
Send D "load & print" (D "load & print"<CRLF>)
Answer: e.g. D A<CRLF>
Send K 3 (K 3<CRLF>)
User
Answer: e.g. K A (K A<CRLF>)
Put weight on
pan and press
print key
Print key
Event: e.g. K 7<CRLF>
Start test
Send K 2 (K 2<CRLF>)
Answer: e.g. K A<CRLF>
Send DW (DW<CRLF>)
Answer: e.g. DW A<CRLF>
Send WS 0 (WS 0<CRLF>)
Answer: e.g. WS A<CRLF>
If answer=WS A 0 then go forward, else send WS again
Send WS (WS<CRLF>)
Answer: e.g. S S 0.00 mg<CRLF>
Create FirstWeight (empty pan)
Send WS 2 (WS 2<CRLF>)
Answer: e.g. WS A<CRLF>
Loop start
Answer: e.g. WS A 0<CRLF>
Send SU (SU<CRLF>)