Operating instructions
Unpacking and setting up the balance
6
Positioning plate of the weighing cell in the desired location and secure (with screws or
by gluing). The foot of the weighing cell should be in the hole (A).
Weighing cell in the positioning plate. Make sure that the front foot of the weighing cell
is in the circular hole. The weighing cell must sit firmly between the two white plastic
bushings (B). For precise positioning, unfasten the fixing screws of the plastic bushings,
move both bushings towards the weighing cell and tighten the screws again.
Remove the shipping lock: Tilt the locking rocker (C) of the weighing chamber cover to
the right and remove the weighing chamber cover by pulling it forwards. Carefully remove
the shipping lock (foam pad with adhesive tape), then replace the weighing chamber
cover and lock.
Note: The shipping lock protects the weighing cell and ensures that the weighing pan
and the weighing sample holder remain fixed in the desired position. Therefore retain the
shipping lock for future transport of the weighing cell.
Level the weighing cell: Turn the two leveling screws (D) at the back of the housing until
the air bubble is in the center of the level indicator (E).
2.4 Assembly of the balance and connection to the power supply
Use the supplied connection cable to connect together the control unit and the weighing cell.
A
B
D
E
C