Operating instructions
Unpacking and setting up the balance
8
Insert the electrodes as far as they will go into the fastening strip (D) and tighten the two
screws (E).
Place the protective cover over the weighing cell.
Connect the electrodes to the power supply unit and the power supply unit to the power
supply (see instructions for the power supply unit).
2.6 Instructions for transporting the balance
When transporting your balance, please observe the corresponding instructions in Chapter 2.7 of the operating instructions for AX and
MX/UMX balances.
When transporting the balance make sure that the weighing sample holder is in the weighing position (lowest position, see Chapter 3).
Then insert the shipping lock (foam pad) in the weighing chamber and close the weighing chamber cover. The shipping lock protects
the weighing cell and ensures that the items in the weighing chamber (weighing pan and weighing sample holder) remain fixed in the
desired position. Secure the weighing chamber cover with a piece of adhesive tape.