8.4 Selection of the adjustment weight
In the model KERN 440, the adjustment weight can be selected from four pre-set
nominal values (approx.1/4; 1/2; 3/4; max) (refer also to table 1 below, factory setting
with grey background). In order to achieve high-quality weighing results in the sense
of the measuring technology, it is recommended to select the nominal value as high
as possible. The non delivered adjustment weights can be purchased from KERN as
option.
Table 1:
440-21N 440-21A 440-33N 440-35N 440-35A 440-43N 440-45N
10g 10g 50g 100g 100g 100g 200g
20g 20g 100g 200g 200g 200g 500g
30g 40g 150g 300g 400g 300g 700g
40g 50g 200g 400g 500g 400g 1,000g
60g 600g
440-47N 440-49N 440-49A 440-51N 440-53N 440-55N
500g 1,000g 1,000g 1,000g 1,000g
1,000g
1,000g 2,000g 2,000g 2,000g 2,000g
2,000g
1,500g 3,000g 5,000g 3,000g 5,000g
5,000g
2,000g 4,000g 6,000g 4,000g 6,000g
6,000g