Data Sheet

HIH-4030/31 Series
2 www.honeywell.com/sensing
TABLE 1. PERFORMANCE SPECIFICATIONS (At 5 Vdc supply and 25 ºC [77 ºF] unless otherwise noted.)
Parameter Minimum Typical Maximum Unit Specific
Note
Interchangeability (first order curve)
0% RH to 59% RH -5
5 % RH
60% RH to 100% RH -8
8 % RH
Accuracy (best fit straight line) -3.5 +3.5 % RH
1
Hysterisis
3
% RH
Repeatability
±0.5
% RH
Settling time
70 ms
Response time (1/e in slow moving air)
5
s
Stability (at 50% RH in a year)
±1.2
% RH
2
Stability (at 50% RH in a year)
±0.5
% RH
3
Voltage supply 4
5.8 Vdc
4
Current supply
200 500 μA
Voltage output (1
st
order curve fit) V
OUT
=(V
SUPPLY
)(0.0062(sensor RH) + 0.16), typical at 25 ºC
Temperature compensation True RH = (Sensor RH)/(1.0546 – 0.00216T), T in ºC
Output voltage temp. coefficient at 50% RH, 5 V
-4
mV/ºC
Operating temperature -40[-40] See Figure 1. 85[185] ºC[ºF]
Operating humidity (HIH-4030) 0 See Figure 1. 100 % RH
5
Operating humidity (HIH-4031) 0 See Figure 1. 100 % RH
Storage temperature -50[-58]
125[257] ºC[ºF]
Storage humidity See Figure 2. % RH
5
Specific Notes:
1. Can only be achieved with the supplied slope and offset. For
HIH-4030/31-003 catalog listings only.
2. Includes testing outside of recommended operating zone.
3. Includes testing for recommended operating zone only.
4. Device is calibrated at 5 Vdc and 25 ºC.
5. Non-condensing environment. When liquid water falls on the
humidity sensor die, output goes to a low rail condition
indicating no humidity.
General Notes:
Sensor is ratiometric to supply voltage.
Extended exposure to >
90% RH causes a reversible
shift of 3% RH.
Sensor is light sensitive. For best performance, shield
sensor from bright light.
FACTORY CALIBRATION DATA
HIH-4030/31 Sensors may be ordered with a calibration and
data printout. See Table 2 and the order guide on the back
page.
TABLE 2. EXAMPLE DATA PRINTOUT
Model HIH-4030-003
Channel 92
Wafer 030996M
MRP 337313
Calculated values at 5 V
V
OUT
at 0% RH
V
OUT
at 75.3% RH
0.958 V
3.268 V
Linear output for 3.5% RH
accuracy at 25 °C
Zero offset
Slope
Sensor RH
0.958 V
30.680 mV/%RH
(V
OUT
- zero offset)/slope
(V
OUT
- 0.958)/0.0307
Ratiometric response for
0% RH to 100% RH
V
OUT
V
SUPPLY