User manual
 56 
UNI-T
®
UT4000 User Manual 
Figure 3-5 Signal noise suppressed 
d  You can also lower waveform brightness to reduce display noise. 
Caution : In the average acquisition mode the waveform display will update at a 
slower rate. This is normal. 
Illustration 5 : Using cursors for measurement 
Your oscilloscope can measure 27 waveform parameters automatically. All auto 
parameters can be measured with the cursors. By using cursors, you can quickly measure 
the time and voltage of a waveform. 
Measuring the frequency of Sinc first peak 
To measure the Sinc frequency of the signal rising edge, follow the steps below : 
1.  Press [CURSOR] to display the cursor measurement menu. 
2.  Press [F1] to activate the cursor measurement function. 
3.  Press [F1] again to set cursor type to time. 
4.  Press [F3] to set the vertical cursor unit to Hz. 
5.  Turn the multi-function control knob to set cursor 1 at the Sinc first peak. 
6.  Press [SELECT] to select cursor 2, then turn the multifunction control knob again to 
set cursor 2 at the Sinc second peak. 
See the figure below. 
Note : When using the cursor to measure voltage, follow step 3 only and set the 
cursor type to voltage. 