UT205E/UT206B UT207B/UT208B 1000A True RMS Digital Clamp Meter P/N:110401109411X
UT205E/UT206B/UT207B/UT208B User Manual Preface Thank you for purchasing the new clamp meter. In order to use this product safely and correctly, please read this manual thoroughly, especially the Safety Instructions part. After reading this manual, it is recommended to keep the manual at an easily accessible place, preferably close to the device, for future reference.
UT205E/UT206B/UT207B/UT208B User Manual UT205E/UT206B/UT207B/UT208B User Manual I. Overview Table of Contents I. Overview ------------------------------------------------------------------- 4 II. Accessories --------------------------------------------------------------- 4 III. Safety Instructions ----------------------------------------------------- 5 IV. Electrical Symbols ----------------------------------------------------- 6 V.
UT205E/UT206B/UT207B/UT208B User Manual III. Safety Instructions The meter is designed and manufactured according to IEC61010-1, IEC610102-032 and IEC61010-2-033 safety standards, and conforms to CAT III 1000V, CAT IV 600V, double insulation, and pollution degree 2. UT205E/UT206B/UT207B/UT208B User Manual IV. Electrical Symbols Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION Note: Before each use, verify meter operation by measuring a known voltage.
UT205E/UT206B/UT207B/UT208B User Manual UT205E/UT206B/UT207B/UT208B User Manual VI. LCD Display (Picture 2, Picture 3, Picture 4) V. External Structure (Picture 1) 8 1. Clamp jaws 2. Hand guard 3. LED indicator light 4. Jaw opening trigger 5. Function dial 6. LCD display 7. Function buttons 8. NCV sensor 9. FLIGHT button 10.Flashlight LED light 11.Battery cover 12.Input jacks 8 1 2 1 3 2 3 9 5 4 4 10 5 6 11 7 6 11 7 12 13 14 Picture 2 UT205E 12 Picture 1 7 9 10 1.
UT205E/UT206B/UT207B/UT208B User Manual UT205E/UT206B/UT207B/UT208B User Manual 10 11 12 13 9 10 11 12 13 14 14 15 16 1 2 3 4 1 2 3 4 5 5 6 7 8 9 6 7 17 8 19 20 21 22 Picture 4 UT206B/UT208B 1. Low battery 15 16 17 18 19 Picture 3 UT207B 9 18 1. Low battery 11. Continuity test 2. Inrush current measurement 12. NCV 3. High voltage 13. Low impedance measurement 4. Auto power off 14. Unit 5. Relative value 15. Analog bar 6. AC signal 16. Auto range 7. DC signal 17.
UT205E/UT206B/UT207B/UT208B User Manual UT205E/UT206B/UT207B/UT208B User Manual VII. Function Dial and Function Buttons 2. Function Buttons 1.Function Dial Description Note: Short press: pressing a button for less than 2s. Long press: pressing a button for more than 2s. 1) SELECT Button Dial Position 位 / / AC/DC current measurement / / AC/DC voltage measurement Ω Resistance measurement Diode test 2) HOLD/¤ Button Continuity test Short press: turn on/off data hold.
UT205E/UT206B/UT207B/UT208B User Manual 6) Hz/INRUSH Button Short press: enter/exit the frequency measurement mode. Only valid for ACV, LPF ACV, LoZ V~, ACA and measurement by flexible current probe. Long press: user has the option to select proper range with RANGE button first, or simply long press this button to enter the inrush current measurement mode (measurement time~100ms). Long press this button again to exit the inrush current measurement mode.
UT205E/UT206B/UT207B/UT208B User Manual 1) AC Current ( ) 2) Inrush Current ( Range Resolution Accuracy 60.00A 0.01A ± (2.0%+5), for UT205E/UT206B ± (2.0%+9), for UT207B/UT208B 600.0A 0.1A 1000A 1A Overload Protection 1000V DC/AC ± (2.0%+5) Display: True RMS Accuracy guarantee: 5%~100% of range. Open circuit allows least significant digit ≤10.
UT205E/UT206B/UT207B/UT208B User Manual 4) AC Voltage ( ) Range Resolution 6.000V 0.001V 60.00V 0.01V 600.0V 0.1V 1000V 1V Accuracy ± (1.2%+3) Overload Protection 1000A ± (1.0%+8) Display: True RMS Accuracy guarantee: 5%~100% of range. Short circuit allows least significant digit ≤5. Input impedance: ≥10MΩ Frequency response: 40Hz~400Hz The AC crest factor will reach 3.0 at 3000 counts while only reach ≤1.5 at 6000 counts.
UT205E/UT206B/UT207B/UT208B User Manual 7) DC voltage ( Range 9) Continuity ( ) Resolution Accuracy 600.0mV 0.1mV ± (0.8%+3) 6.000V 0.001V 60.00V 0.01V 600.0V 0.1V 1000V 1V ± (0.5%+5) Overload Protection 1000A Input impedance: ≥10MΩ Accuracy guarantee: 5%~100% of range. Short circuit allows least significant digit ≤5. Resolution Accuracy ± (1.0%+3) 600.0Ω 0.1Ω 6.000kΩ 0.001kΩ 60.00kΩ 0.01kΩ 600.0kΩ 0.1kΩ 6.000MΩ 0.001MΩ 60.00MΩ 0.01MΩ ± (1.0%+2) ) Resolution Accuracy 0.
UT205E/UT206B/UT207B/UT208B User Manual Measurement result = displayed value – capacitance of open-circuit test leads For capacitance ≤1μF, it is recommended to use “REL” measurement mode. Accuracy guarantee: 5%~100% of range 12) Temperature (°C/°F) Range Resolution Accuracy -40℃~300℃ 0.1℃ ±(1.0%+20) 300℃~1000℃ 1℃ ±(1.0%+2) -40℉~572℉ 0.2℉ ±(1.0%+40) 573℉~1832℉ 1℉ ±(1.0%+4) Overload Protection 1000V Only K-type thermocouple is applicable.
UT205E/UT206B/UT207B/UT208B User Manual AC Current Measurement 1) Turn the function dial to A~, , or position. 2) Press the trigger to open clamp jaws and fully enclose one conductor (only one conductor can be measured at a time). For optimum results, center the conductor in the jaws. Current Frequency Measurement 1) When the function dial is in the AC current position, short press the Hz or Hz/ INRUSH button to enter the frequency measurement mode.
UT205E/UT206B/UT207B/UT208B User Manual 3. Measurement by Flexible Current Probe (Picture 6) UT205E/UT206B/UT207B/UT208B User Manual Inrush Current Measurement 1) After connecting the flexible current probe, short press the RANGE button to select the proper range. 2) Long press the Hz/INRUSH button to enter the inrush current measurement mode. 3) Start the appliance under test and measure the instantaneous inrush current of the appliance.
UT205E/UT206B/UT207B/UT208B User Manual 4. Related Measurement of AC Voltage and LPF ACV (Picture 7) UT205E/UT206B/UT207B/UT208B User Manual Voltage Frequency Measurement 1) When the function dial is in the AC voltage, short press Hz or Hz/INRUSH button to enter the frequency measurement mode. 2) Short press Hz or Hz/INRUSH button again to exit frequency measurement mode. LPF ACV Measurement 1) When the function dial is in the AC voltage position, long press SELECT button to enable LPF ACV function.
UT205E/UT206B/UT207B/UT208B User Manual 5. DC Voltage Measurement (Picture 7) 1) Insert the red test lead into the , or jack, and black test lead into the COM jack. 2) Turn the function dial to the or position. 3) Short press the SELECT button to switch to DC voltage measurement if required, and connect the test leads with the measured load or power supply in parallel. 4) Read the voltage value on the display.
UT205E/UT206B/UT207B/UT208B User Manual 7. Resistance Measurement (Picture 9) UT205E/UT206B/UT207B/UT208B User Manual When measuring resistance above 1MΩ, it is normal to take a few seconds to stablize reading. Use caution when working with voltage above AC 30Vrms, 42Vpeak or DC 60V. Those voltages may pose shock hazard. After completing the measurement, disconnect the test leads from the circuit under test. 8.
UT205E/UT206B/UT207B/UT208B User Manual 3) Connect the red probe with the diode anode, and black with the diode cathode. 4) Read the approximate forward voltage of the diode on the display. For silicon PN junction, the normal value is generally about 500~800 mV. Note: If the diode is open or its polarity is reversed, the LCD will display “OL”. Before measuring the diode online, switch off the power supply of the circuit, and fully discharge all capacitors.
UT205E/UT206B/UT207B/UT208B User Manual 1) Insert the red test lead into the , or jack, and black into the COM jack. 2) Turn the function dial to the Hz% position, and connect the test leads with both ends of the measured signal source in parallel. 3) Short press the SELECT button to switch to frequency/duty ratio measurement. Note: Do not input voltage higher than 30Vrms to avoid personal injury. After completing the measurement, disconnect the test leads from the circuit under test. 12.
UT205E/UT206B/UT207B/UT208B User Manual 1) Turn the function dial to the NCV position, and bring the NCV sensor close to the wire under test. 2) If there is AC voltage or electromagnetic field in the space, the LCD will display the sensing intensity from weak to strong by “—”. At the same time, the buzzer will beep intermittently and the LED indicator light will be on. When no voltage is sensed, the LCD displays “EF”. 14.
UT205E/UT206B/UT207B/UT208B User Manual 39 UT205E/UT206B/UT207B/UT208B User Manual 40