User Manual

13 14
VIII. Specifications
1. General Specifications
Max display: ------------------------- 6000 counts
Polarity display: --------------------- Auto
Overload display: ------------------ “OL” or “-OL”
Low battery indication: ------------ “ ” is displayed.
Sampling rate: ---------------------- 3 times/s
Sensor type: ------------------------- Coil induction (UT205E/UT206B)
Hall effect sensor (UT207B/UT208B)
Test position error: ---------------- If the source under test is not placed at the
center of the clamp jaws when measuring current,
±1.0% additional error in reading may be produced.
Jaw opening: ------------------------ 42mm
Battery: ------------------------------- 3×1.5V AAA
Auto power off: ---------------------- 15 minutes (can be disabled)
Dimensions: -------------------------- 272mm×81mm×43.5mm
Weight (including batteries): ----- About 492g (UT205E/UT206B),
447g (UT207B/UT208B)
2. Environmental Specifications
Operating altitude:----------------------------- 2000m
Safety standards:----------------------------- IEC61010-1, IEC61010-2-032,
IEC61010-2-033;
CAT III 1000V, CAT IV 600V
Pollution degree:------------------------------- 2
Operating temperature and humidity:----- 0°C~30°C (80%RH), 30°C~40°C
(75%RH), 40°C~50°C (45%RH)
Storage temperature and humidity:------ -10°C~60°C (80%RH)
Electromagnetic compatibility:-------------- When RF=1V/m:
overall accuracy = specified accuracy
+ 5% of range
When RF>1V/m: not specified
3. Electrical specifications
Accuracy:--------------------------------------- ± (a% of reading + b digits),
1 year calibration cycle
Ambient temperature:----------------------- 23°C±5°C
Ambient humidity:---------------------------- 80%RH
Note:
To ensure measurement accuracy, the operating temperature should be within
18°C~28°C and the fluctuation range should be within ±1°C. When the temperature
is <18°C or >28°C, add temperature coefficient error 0.1 x (specified accuracy)C.
UT205E/UT206B/UT207B/UT208B User Manual UT205E/UT206B/UT207B/UT208B User Manual
6) Hz/INRUSH Button
Short press: enter/exit the frequency measurement mode.
Only valid for ACV, LPF ACV, LoZ V~, ACA and measurement by flexible current
probe.
Long press: user has the option to select proper range with RANGE button first,
or simply long press this button to enter the inrush current measurement mode
(measurement time~100ms). Long press this button again to exit the inrush
current measurement mode.
Inrush current can also be measured with flexible current probe (UT206B/UT208B).
7) Hz Button (UT205E only)
Short press: enter/exit the frequency measurement mode.
8) FLIGHT Button
Short press: turn on/off the flashlight.