Specifications
15
Measurement Operation (3)
C. Measuring Resistance (See figure 5)
   Warning
To avoid damages to the Meter or to the devices under test, 
disconnect circuit power and discharge all the high-voltage 
capacitors before measuring resistance.
The resistance ranges are: 400.0Ω , 4.000kΩ , 40.00kΩ , 400.0kΩ , 
4.000MΩ and 40.00MΩ . To measure resistance, connect the Meter 
as follows:
1. Insert the red test lead into the 
HzΩ terminal and the black 
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to Ω.
Model UT206: Set the rotary switch to 
Ω , resistance 
measurement (Ω) is default or press BLUE button to select Ω 
measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
 • The test leads can add 0.1Ω to 0.2Ω of error to resistance 
measurement. To obtain precision readings in low-resistance 
measurement, that is the range of 400.0Ω , short-circuit the input 
terminals beforehand, using the relative measurement function button 
REL
∆
 to automatically subtract the value measured when the 
testing leads are short-circuited from the reading.
 • If Ω reading with shorted test leads is not ≤0.5Ω , check for 
loose test leads, incorrect function selection, or enabled Data Hold 
function.
 • For high-resistance measurement (>1MΩ ), it is normal to take 
several seconds to obtain a stable reading.
 • The LCD displays 
 indicating open-circuit for the tested resistor 
or the resistor value is higher than the maximum range of the Meter.
red
black
red
black
Model UT205/UT206: OPERATING MANUAL