Specifications
Model UT90A: OPERATING MANUAL
10
To avoid damages to the built-in fuse and the
Meter, please do not attempt parallel connect
the test leads with any circuit while the test
leads are inserted into the current terminal.
For the safety requirement, when measure
high current, every measure time should< 10
seconds and the time between two
measurements should > 15 minutes.
The measurement ranges of DC current are:
200mA,2mA,20mA,200mA,20A; The
measurement ranges of AC current are:
200.0µA,2.000mA,20.00mA,200.0mA and
20.00A. To measure current, do the following:
1  Insert the red test lead into the“µA
 mA ”or“20A”terminal and the black test
 lead into the COM terminal.
2  Set the rotary switch to“A ”or“A ”,and
 Connect the test leads across with the
 circuit being measured.
3.  The measured value shows on the display,
 it is an effective value of sine wave (mean
 value response).
Note
l  Turn off the power of circuit being
 measured before series connection
 between the Meter and the circuit.
l  Use proper terminals and ranges for the
 measurement.
l  Please use the ranges from high to low if
 the current not being confirmed.
l  When current measurement has been