Specifications
11
1.  Insert the red test lead into the HzVȍterminal and 
the black test lead into the COM terminal. 
2.  Set the rotary switch toȍ
, resistance 
measurement (ȍ) is default (or press SELECT
button to selectȍmeasurement mode) 
3. Connect the test leads with the object being 
measured. 
  The measured value shows on the display. 
Note
z  The test leads can add 0.1ȍto 0.2ȍ of error to 
resistance measurement. To obtain precision 
readings in low-resistance measurement, that is 
the range of 400.0ȍ, short-circuit the input 
terminals beforehand, using the relative value 
function button 
RESET to automatically 
subtract the value measured when the testing 
leads are short-circuited from the reading. 
z  For high-resistance measurement (>1Mȍ), it is 
normal taking several seconds to obtain a stable 
reading.
z If ȍ reading with shorted test leads is not 0.5ȍ,
check for loose test leads, incorrect function 
selection, or enabled Data Hold function. 
z The LCD displays OL indicating open-circuit for 
the tested resistor or the resistor value is higher 
than the maximum range of the Meter.   
z  When resistance measurement has been 
completed, disconnect the connection between 
the test leads and the circuit under test. 
Testing Diodes (see figure 5) 
Use the diode test to check diodes, transistors, and 
72-9375: OPERATING MANUAL










