Specifications
12
other semiconductor devices.    The diode test sends a 
current through the semiconductor junction, and then 
measures the voltage drop across the junction. A 
good silicon junction drops between 0.5V and 0.8V.
To test a diode out of a circuit, connect the Meter as 
follows: 
1.  Insert the red test lead into the HzVȍterminal and 
the black test lead into the COM terminal. 
2. Set the rotary switch toȍ
and press 
SELECT button to select 
 measurement 
mode.
3.  For forward voltage drop readings on any 
semiconductor component, place the red test 
lead on the component’s anode and place the 
black test lead on the component’s cathode. 
  The measured value shows on the display. 
Note
z  In a circuit, a good diode should still produce a 
forward voltage drop reading of 0.5V to 0.8V; 
however, the reverse voltage drop reading can 
vary depending on the resistance of other 
pathways between the probe tips. 
z  Connect the test leads to the proper terminals as 
listed above.    The LCD will display OL indicating 
open-circuit for wrong connection. The unit of 
diode is Volt (V), displaying the 
positive-connection voltage-drop value. 
z  When diode testing has been completed, 
disconnect the connection between the testing 
leads and the circuit under test. 
Continuity Testing (see figure 6) 
72-9375: OPERATING MANUAL










