User Guide
19
Measurement Operation(6)
red black
( figure 6)
Model UT90C: OPERATING MANUAL
l Connect the test leads to the proper terminals as
 said above to avoid error display. The LCD will
 display OL indicating open-circuit for wrong connection.
 The unit of diode is Volt (V), displaying the positive-
 connection voltage-drop value.
l When diode testing has been completed, disconnect
 the connection between the testing leads and the
 circuit under test.
Continuity Measurement (see figure 6)
To test for continuity, connect the Meter as below:
1. Insert the red test lead into the HzVΩ terminal and
 the black test lead into the COM terminal.
2. Set the rotary switch toΩ and press SELECT
 button to select measurement mode.
3. The buzzer does not sound if the circuit is disconnected
 with resistance value is > 100Ω
 The buzzer sounds continuously if the circuit is in
 good condition with resistance value 10Ω.
4. The nearest circuit resistance value shows on the
 display, the unit isΩ.
Note
l The LCD displays OL indicating the resistance of
 the circuit being tested is higher than 400Ω.
l The buzzer sounds once if the RESET, SELECT
 or is pressed on.










