UT203/204 P/N: 41416301 Registered Design No.: ZL02 3 357791.
Model UT203/204: OPERATING MANUAL Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons The Effectiveness of Functional Buttons Display Symbols Measurement Operation A. DC/AC Voltage Measurement B. Measuring Resistance C. Testing Diodes D. Testing for Continuity E. Frequency Measurement F. Duty Cycle Measurement G.
Model UT203/204: OPERATING MANUAL Title Sleep Mode Specifications A. General Specifications B. Environmental Requirements Accuracy Specifications A. DC Voltage B. AC Voltage C. Resistance D. Diode Test E. Continuity Test F. Frequency G. Duty Cycle H. DC Current I. AC Current Maintenance A. General Service B.
Model UT203/204: OPERATING MANUAL Overview This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly. Warning To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safe Operation” carefully before using the Meter.
Model UT203/204: OPERATING MANUAL Unpacking Inspection Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part: Item 1 2 3 4 5 Description English Operating Manual Test Lead Test Clip Carrying Bag 9V Battery (NEDA1604, 6F22 or 006P) Qty 1 piece 1 pair 1 pair 1 piece 1 piece In the event you find any missing or damage, please contact your dealer immediately.
Model UT203/204: OPERATING MANUAL Safety Information This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. II 600V, CAT. III 300V) and double insulation. CAT. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than CAT. III. CAT. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT.
Model UT203/204: OPERATING MANUAL Rules For Safe Operation Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules: l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. l Inspect the test leads for damaged insulation or exposed metal.
Model UT203/204: OPERATING MANUAL l The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter. l Do not carry out the measurement when the Meter’s back case and battery compartment are not closed to avoid electric shock. l Do not input higher than 600V between the two Meter’s input terminal to avoid electric shock and damages to the Meter.
Model UT203/204: OPERATING MANUAL l When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. l The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. l Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident. l The Meter is suitable for indoor use.
Model UT203/204: OPERATING MANUAL International Electrical Symbols AC (Alternating Current) DC (Direct Current) AC or DC Grounding Double Insulated Warning.
Model UT203/204: OPERATING MANUAL The Meter Structure (see figure 1) 1. 2. 3. 4. 5. Input Terminals LCD Display Functional Buttons Rotary Switch Trigger: press the lever to open the transformer jaws. When the pressure on the lever is released, the jaws will close. 6. Hand Guards: to protect user’s hand from touching the dangerous area. 7. Transformer Jaw: designed to pick up the AC and DC current flowing through the conductor. It could transfer current to voltage.
Model UT203/204: OPERATING MANUAL Rotary Switch Below table indicated for information about the rotary switch positions.
Model UT203/204: OPERATING MANUAL Functional Buttons Below table indicated for information about the functional button operations. Button HOLD REL Operation Performed l Press HOLD to enter the Hold mode in any mode, the Meter beeps. l Press HOLD again to exit the Hold mode, the Meter beeps. At V and range: l Press to select manual ranging measurement mode. The Meter is default to auto ranging measurement mode. l When the Meter is at manual ranging measurement mode, press to step down the range.
Model UT203/204: OPERATING MANUAL Button SELECT Operation Performed l Press SELECT button to select the alternate functions marked in blue colour on the Meter’s faceplate including Hz, Duty%, V , , 40 A , and 400 A . l After the Meter entering Sleep Mode, press and hold SELECT to turn the Meter on, it will disable the Sleep Mode feature.
Model UT203/204: OPERATING MANUAL The Effectiveness of Functional Buttons Not every functional buttons can be used on every rotary switch positions.
Model UT203/204: OPERATING MANUAL Display Symbols (see figure 2) 4 7 8 9 15 3 5 6 2 10 14 1 16 (figure 2) 15 11 12 13
Model UT203/204: OPERATING MANUAL Number 1 2 3 Symbol AC DC 4 5 6 7 8 9 % Meaning Indicator for AC voltage or current Indicator for DC voltage The battery is low. Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. The Meter is in the auto range mode in which the Meter automatically selects the range with the best resolution. Test of diode The continuity buzzer is on Indicator for Duty.
Model UT203/204: OPERATING MANUAL Number 10 Symbol ,k ,M 11 12 13 Hz A mV, V 14 15 16 TRMS OL Meaning : Ohm. The unit of resistance. k : Kilohm. 1x103 or 1000 ohms M : Megohm. 1x106 or 1,000,000 ohms The unit of Frequency Amperes (amps). The unit of current. Volts. The unit of voltage. mV: Millivolt. 1x10-3 or 0.
Model UT203/204: OPERATING MANUAL Measurement Operation A. DC/AC Voltage Measurement (see figure 3) Warning To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained. Red The DC Voltage ranges are: 400mV, 4V, 40V, 400V and 600V. The AC Voltage ranges are: 4V, 40V, 400V and 600V. Black (figure 3) To measure DC voltage, connect the Meter as follows: 1.
Model UT203/204: OPERATING MANUAL 2. Set the rotary switch to V . DC mesaurement mode and auto ranging is a default. Press SELECT to switch to AC measurement mode or press REL to switch to manual ranging measurement mode. 3. Connect the test leads across with the object being measured. The measured value shows on the display.
Model UT203/204: OPERATING MANUAL B. Measuring Resistance (see figure 4) Warning To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance. The resistance ranges are: 400 , 4k , 40k ,400k ,M and 40M . Red Black To measure resistance, connect the Meter as follows: 1. Insert the red test lead into the Hz Duty% (figure 4) terminal and the black test lead into the COM terminal. 2.
Model UT203/204: OPERATING MANUAL Note l To obtain a more precise reading, you could remove the objects being tested from the circuit when measuring. l When resistance measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
Model UT203/204: OPERATING MANUAL C. Testing Diodes (see figure 5) Warning To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before testing diodes. Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test sends a current through the semicondutor junction, then measure the voltage drop across the junction. A good silicon junction drops between 0.5V and 0.8V.
Model UT203/204: OPERATING MANUAL 2. Set the rotary switch to . Diode measurement mode is a default or presss SELECT to select measurement mode. 3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode. Note l To obtain a more precise reading, you could remove the objects being tested from the circuit when measuring.
Model UT203/204: OPERATING MANUAL D. Testing for Continuity (see figure 6) Warning To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity. To test for continuity, connect the Meter as follows: Red Black 1. Insert the red test lead into the Hz Duty% terminal and the black test lead into the 2. COM terminal. Set the rotary switch to and press SELECT (figure 6) 3. button to select measurement mode.
Model UT203/204: OPERATING MANUAL Note l When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
Model UT203/204: OPERATING MANUAL E. Frequency Measurement (see figure 7) Warning To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained. The resistance ranges are: 10Hz, 100Hz, 1kHz, 10kHz, 100kHz, 1MHz and 10MHz. Red Black To measure frequency, connect the Meter as follows: 1.
Model UT203/204: OPERATING MANUAL Note l When frequency measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
Model UT203/204: OPERATING MANUAL F. Duty Cycle Measurement (see figure 8) Warning To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained. The duty cycle range is: 0.1%~99.9%. Red Black To measure duty cycle, connect the Meter as follows: 1. Insert the red temperature probe into the Hz Duty% terminal and the black temperature probe (figure 8) into the COM terminal. 2.
Model UT203/204: OPERATING MANUAL Note l When duty cycle measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
Model UT203/204: OPERATING MANUAL G. DC/AC Current Measurement (see figure 9) The measuremnet ranges of current are: 40.00 A and 400.0 A . To measure current, do the following: 1. Set the rotary switch to 40 A or 400 A . DC measuremnet mode is a default. Press SELECT to switch between DC and AC measurement mode. 2. Hold the Meter tight, don’t release. The Hall components are very senstive not only to the magnet but also to heat and machines reaction force.
Model UT203/204: OPERATING MANUAL Note l Press REL to subtracts a stored value from the present measurement value and displays a result. l When current measurement has been completed, disconnect the connection between the conductor under test and the jaw, and remove the conductor away from the transformer jaw of the Meter.
Model UT203/204: OPERATING MANUAL Sleep Mode To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 15 minutes. The Meter can be activated by turning the rotary switch or pressing the button based on ‘The Effectiveness of Functional Buttons” on page 14. Press SELECT to activate the Meter will disable the Sleep Mode feature.
Model UT203/204: OPERATING MANUAL Specifications A. General Specifications: l Maximum Voltage between any Terminals and grounding: Refer to different range input protection voltage. l Display: 3 3/4 digits LCD display, Maximum display 3999 l Polarity: Automatically display. l Overloading: Display OL or –OL l Battery Deficiency: Display l Measurement Speed: Updates 3 times/second.
Model UT203/204: OPERATING MANUAL l Battery Life: typically 150hours (alkaline battery) l Sleep Mode (can be disabled) l Dimensions (H x W x L): 208mm x 76mm x 30mm. l Weight: Approximate 260g (battery included) B. Environmental Requirements l The Meter is suitable for indoor use. l Altitude: Operating: 2000m Storage: 10000m l Safety/ Compliances: IEC 61010 CAT.II 600V, CAT.III 300V over voltage and double insulation standard. l Temperature and humidity: Operating: 0ºC~30ºC ( 85%R.H); 30ºC~40ºC ( 75%R.
Model UT203/204: OPERATING MANUAL Accurate Specifications Accuracy: (a% reading + b digits), guarantee for 1 year. Operating temperature: 23ºC 5ºC Relative humidity: 85%R.H Temperature coefficient: 0.1x(specified accuracy)/1ºC A. DC Voltage Range 400.0mV 4.000V 40.00V 400.0V 600V Resolution 0.1mV 1mV 10mV 100mV 1V Accuracy (0.8%+3) Overload protection (0.
Model UT203/204: OPERATING MANUAL B. AC Voltage Range 4.000V 40.00V 400.0V 600V Resolution 1mV 10mV 100mV 1V Accuracy Overload protection (1%+5) 600V DC/AC (1.2%+5) Remarks: l Input impedance: 10M // less than 100pF l Frequency response: 40Hz~400Hz. l Change to AC: UT203: Change to AC by using average response method. Input sine wave, then adjust the reading until it is same as the effective value. UT203: Combine AC and True RMS response method. Input sine wave to adjust.
Model UT203/204: OPERATING MANUAL C. Resistance Range 400.0 4.000k 40.00k 400.0k 4.000M 40.00M Resolution 100m 1 10 100 1k 10k Accuracy (1.2%+2) (1%+2) Overload protection 600Vp (1.2%+2) (1.5%+2) D. Diode Test Range Resolution 1mV Accuracy Overload protection Display forward voltage 600Vp drop nearest value Remark: Open circuit voltage approximate 1.48V.
Model UT203/204: OPERATING MANUAL E. Continuity Test Range Resolution 100m Accuracy Around 50 , the buzzer beeps Overload protection 600Vp Remark: l Open circuit voltage approximate 0.45V. l The buzzer may or may not beeps when the resistance of a circuit under test is between 50 ~100 l The buzzer will not beep when the resistance of a circuit under test is > 100 .
Model UT203/204: OPERATING MANUAL F. Frequency Range 10Hz 100Hz 1kHz 10kHz 100kHz 1MHz 10MHz Resolution 0.001Hz 0.01Hz 0.1Hz 1Hz 10Hz 100Hz 1kHz Accuracy Overload protection (0.
Model UT203/204: OPERATING MANUAL G. Duty Cycle Range Resolution Accuracy Overload protection 0.1%~99.9% 0.1% For reference only 600Vp Resolution 0.01A 0.1A Accuracy (2%+5) (2%+3) Overload protection H. DC Current Range 40.00A 400.0A 400A DC/AC Warning The operating temperature must be 0ºC ~40ºC when measuring current. Remark: l If the reading is positive, the current direction is from bottom to up. See figure 10, the front case face up while the bottom case face down.
Model UT203/204: OPERATING MANUAL also to heat and machines reaction force. Any shock will cause the changing in reading in the short time. Follow the below procedure to measure current will be more precise: l Hold the Meter tight and press the lever to open the transformer jaw. Center the conductor within the transformer jaws, then release the Meter slowly until the transformer jaw is completely closed.
Model UT203/204: OPERATING MANUAL I. AC Current Range 40.00A 400.0A Resolution Accuracy Frequency Response Overload protection 0.01A (2.5%+8) 400A DC/AC 50Hz ~ 60Hz 0.1A (2.5%+5) Warning The operating temperature must be 0ºC ~40ºC when measuring current. Remark: l It may have 10 digits or less unstable or wrong digits, it will not affect measurement result. l Hold the Meter tight, do now release. The Hall components are very sensitive not only to the magnet but also to heat and machines reaction force.
Model UT203/204: OPERATING MANUAL 2. 3. 4. 5. Remove the transformer jaw. Press REL to display zero. Repeat the above 1. procedure. The obtained reading will be more precise. l Change to AC: UT203: Change to AC by using average response method. Input sine wave, then adjust the reading until it is same as the effective value. UT204: Combine AC and True RMS response method. Input sine wave to adjust. Non sine wave must follow the below data to adjust: Peak factor: 1.4~2.0, add 1.
Model UT203/204: OPERATING MANUAL MAINTENANCE This section provides basic maintenance information including battery replacement instruction. Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case. A. General Service l Periodically wipe the case with a damp cloth and mild detergent.
Model UT203/204: OPERATING MANUAL B. Replacing the Battery (see figure 10) Warning Screw To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. Make sure the transformer jaw and the tets leads are disconected from the circuit being tested before opening the case bottom. To replace the battery: (figure 10) 1. Turn the Meter off and remove all the connections from the input terminals 2.
Model UT203/204: OPERATING MANUAL Copyright 2005 Uni-Trend Group Limited. All rights reserved. Manufacturer: Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925 Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.