Model UT601: OPERATING MANUAL Table of Contents Title Page Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Functional Buttons Display Symbols Measurement Operation A. Measuring Resistance B. Capacitance Measurement C. Diode and Continuity Test D. Transistor hFE Measurement General Specifications Accuracy Specifications A. Resistance Test B. Capacitance Test C. Continuity & Diodes D. Transistor Maintenance A. General Service B.
Model UT601: OPERATING MANUAL 2
Model UT601: OPERATING MANUAL Overview This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly. Warning To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safety Operation” carefully before using the Meter.
Model UT601: OPERATING MANUAL Unpacking Inspection Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part: Item 1 Description English Operating Manual Qty 1 piece 2 Test Clip 1 pair 3 9V Battery (NEDA1604, 6F22 or 006P) 1 piece (installed) In the event you find any missing or damage, please contact your dealer immediately.
Model UT601: OPERATING MANUAL Safety Information This Meter complies with the standards EMC EN61326. Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired. In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test. A Note identifies the information that user should pay attention on.
Model UT601: OPERATING MANUAL Rules For Safe Operation Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules: l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. l Inspect the test clips for damaged insulation or exposed metal.
Model UT601: OPERATING MANUAL l When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. l The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. l Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident. l The Meter is suitable for indoor use.
Model UT601: OPERATING MANUAL International Electrical Symbols Grounding Double Insulated Deficiency of Built-In Battery. Continuity Test. Diode. Capacitance Test Fuse. Warning. Refer to the Operating Manual. Conforms to Standards of European Union.
Model UT601: OPERATING MANUAL The Meter Structure (see figure 1) 1 2 8 7 3 6 5 4 (figure 1) 1. LCD Display 2. Capacitance Zero Adjustment Switch 3. Transistor Jack 4. Resistance, Diode and Continuity Buzzer Input Terminal 5. Capacitance Input Terminal 6. Small Value Capacitance Jack 7. Rotary Switch . 8. Power Button Functional Buttons Below table indicated for information about the functional button operations. Button Power Description Press the Power down to turn the Meter on.
Model UT601: OPERATING MANUAL Display Symbols (see figure 2) 3 1 mF µF nF pF mH ΩH kΩMΩ 2 5 4 6 8 7 (figure 2) No. Symbol Meaning 1 Data hold is active. 2 The battery is low. Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. 3 β The continuity buzzer is on. 5 6 Transistor Test Test of diode. 4 pF, nF, Farad. The unit of capacitance F, mF pF: Picofarad. 1x10-12 or 0.
Model UT601: OPERATING MANUAL Measurement Operation is not on, l Make sure the Low Battery Display otherwise false readings may be provided. symbol, before carrying l Pay extra attention to the measurement, which is located besides the input terminals of the Meter. A. Measuring Resistance (see figure 3) (figure 3) Warning To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance.
Model UT601: OPERATING MANUAL Note l When measuring at 20Ω and 200Ω range, the test clips can add 0.1 to 0.3Ω error to resistance. To obtain precise readings in these low-resistance measurement, that is the range 20Ω and 200Ω, short circuit the input terminals beforehand and record the reading obtained (called this reading as X). (X) is the additional resistance from the test clips. Then use the equation: measured resistance value (Y) – (X) = precision readings of resistance.
Model UT601: OPERATING MANUAL B. Capacitance Measurement (see figure 4) (figure 4) Warning To avoid damage to the Meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC Voltage function to confirm that the capacitor is discharged. The Meter’s capacitance ranges are:200pF, 2nF, 20nF, 200nF, 2µF, 20µF, 200µF, 2mF and 20mF. To measure capacitance, connect the Meter as follows: 1.
Model UT601: OPERATING MANUAL 4. When measuring small value capacitor, that is 200pF, 2nF and 20nF, first open circuit the test clips or the Small Value Capacitance Jack, then turn the Capacitance Zero Adjustment Switch to adjust zero. 5. The measured value shows on the display. Note l Do not short the test clips to avoid the consumption of battery.
Model UT601: OPERATING MANUAL C. Diode and Continuity Test (see figure 5) red (figure 5) Warning To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring diodes and continuity. Testing Diodes Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test sends a current through the semiconductor junction, and then measures the voltage drop across the junction.
Model UT601: OPERATING MANUAL Note l In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however, the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips. l Connect the test clips to the proper terminals as said above to avoid error display. The LCD will display “1” indicating open-circuit for wrong connection. The unit of diode is Volt (V), displaying the positive-connection voltage-drop value.
Model UT601: OPERATING MANUAL D. Transistor hFE Measurement (see figure 6) NPN PNP hFE (figure 6) To measure transistor, set up the Meter as follows: 1. Check that the transistor is PNP or NPN type. 2. Insert the transistor to be measured to the corresponding Transistor Jack. 3. The Meter displays the tested transistor’s nearest value.
Model UT601: OPERATING MANUAL General Specifications l Fused Protection for capacitance Input Terminal: 0.315A, 250V, fast type fuse, 5x20 mm. l Maximum Display: Display: 1999. l Measurement Speed: Updates 2-3 times /second. l Polarity: Auto. (Display “-“ when negative) Display “1” l Overloading: l Range: Manual Ranging l Capacitance range zero adjustment: The range around 20pF l Temperature: o o o o Operating: 0 C~40 C (32 F ~104 F). o o o o Storage: -10 C~50 C (14 F ~122 F).
Model UT601: OPERATING MANUAL Accuracy Specifications Accuracy: (a% reading + b digits), guarantee for 1 year. o o Operating temperature: 23 C 5 C. Relative humidity: <75%. o Temperature coefficient: 0.1 x (specified accuracy) / 1 C. A. Resistance Test Range Resolution Accuracy 20Ω 0.01Ω (1%+5) 200Ω 0.1Ω (0.8%+3) 2kΩ 0.001kΩ 20kΩ 0.01kΩ 200kΩ 0.1kΩ Overload Protection (0.8%+1) 250V rms 2MΩ 0.001MΩ 20MΩ 0.01MΩ (1%+2) 200MΩ 0.
Model UT601: OPERATING MANUAL B. Capacitance Test Range Resolution 200.0pF 0.1pF 2.000nF 0.001nF 20.00nF 0.01nF 200.0nF 0.1nF 2.000µF 0.001µF 20.00µF 0.01µF 200.0µF 0.1µF 2.000mF 0.001mF 20.00mF 0.01mF Accuracy Testing Frequency 800Hz (0.5%+10) 80Hz (2%+2) 8Hz Remarks: l Overload Protection 0.315A, 250V, fast type fuse, 5x20 mm l If the Meter can not adjust to zero, you could use the tested values minus the open circuit value to get the correct measurement value.
Model UT601: OPERATING MANUAL D. Transistor Range Resolution hFE 1β Testing Overload Condition Protection Vce 2.8V I bo 10µA 250V rms Remarks: l The display value is the tested transistor’s nearest value (0~1000β).
Model UT601: OPERATING MANUAL Maintenance This section provides basic maintenance information including battery and fuse replacement instruction. Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case. A. General Service l Periodically wipe the case with a damp cloth and mild detergent.
Model UT601: OPERATING MANUAL B. Replacing the Battery (see figure 7) (figure 7) Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. To replace the battery: 1. Turn the Meter power off and remove all connections from the terminals. 2. Remove the screw from the battery compartment, and separate the battery compartment from the case bottom. 3. Remove the battery from the battery compartment. 4.
Model UT601: OPERATING MANUAL C. Replacing the Fuse (see figure 8) SC RE W (figure 8) Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure. To replace the Meter’s fuse: 1. Turn the Meter power off and remove all connections from the terminals. 2. Remove the screw from the battery compartment, and separate the battery compartment from the case bottom. 3.
Model UT601: OPERATING MANUAL Replacement of the fuses is seldom required. Burning of a fuse always results from improper operation.
Model UT601: OPERATING MANUAL Copyright 2005 Uni-Trend Group Limited. All rights reserved. Manufacturer: Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925 Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.