User's Guide

USER GUIDE: UBS-200L LF DESKTOP READER
April 4, 2019
Unified Information Devices | Appendix 3 Connecting to Third-Party Systems
50
2. The “Generic Interface Setup” Screen appears.
3. Populate the fields according to the values the table below.
Interface ID:
(Use Something recognizable, such as UI Devices UBS-200L)
Communication Parameters
Serial Port:
COMX (Where X is the COM port value determined in Chapter 2
Getting Started with the UBS-200L
Baud Rate:
38400
Parity:
None
Word Length:
8
Stop Bits:
1
Instrument Output Parameters
Output Data Length:
9
Result Start:
1
Result End:
8
Data Terminator:
<CR>
Initialisation Cmd:
None
Acknowledge Send:
None
Acknowledge Receive:
None
4. Select “Test” at the bottom of the Setup screen.
5. Select “Implant Reader” from the “Interface Type” box.
6. Present a UI Devices transponder to the UBS-200L antenna within 5 seconds. The 8-character
Transponder ID should display perfectly just below the text “Last string received from the
instrument” and should be formatted correctly in the “Result Value” box.
7. Set-up is complete! Scanned Transponder IDs will now appear in the “Implant Number” field of
the “Select Subject” window.