User's Guide

36
Warning/Error
Screen
Cause
Solution
70-22
70-22
ERROR
Analyzer tilted
The analyzer must not be
tilted or moved during
the test. Restart the test
with a new strip.
During sample application,
the analyzer was tilted too
much or was bumped.
Discard the strip before
starting a new test. Do not
bump or tilt the analyzer after
the sample is applied.
70-24
70-24
ERROR
Battery critical
The battery must be
charged. Connect the
analyzer to power before
starting a new test.
The battery is too low to start
a new test.
Before a new test can be
started you must connect the
analyzer to an external power
source, which will also charge
the battery.
17-0000
17-0000
ERROR
Battery too low
Connect to external
power and wait
for the battery
to charge enough
The battery is too low for the
analyzer to operate, most
likely because the analyzer
has not been used for a very
long time.
Before the analyzer can be
used, you must connect the
analyzer to an external power
source to charge the battery
enough for it to be used
safely. This may take 30
minutes or more.