Use Instructions

Table Of Contents
48
Caution
1. The check strip only used for daily check.
2. Please do not soak the check strip into any
liquid.
3. Please keep the check strip clean.
4. Compare the obtained results with the range
printed on the check strip container. If the
obtained results inside the range that means
the instrument can be normal use; if not please
check the check strip is all right or not.
5. The range printed on the check strip
container can only be used as the basis for
judging instruments are in good condition, not
as a reference for clinical diagnosis.
5.2 Checkup Operate
1) Insert the strip tray into the tray conveyor of the
instrument. Note that the handle end of the reagent
strip should be on the outside of the instrument,
and make sure that the end of the strip tray touches