+ Due to the high compression ratio, the pump may gen-
erate overpressure at the outlet. Check pressure com-
patibility with system components (e.g., exhaust pipe-
line or exhaust valve) at the outlet.
+ Do not permit any uncontrolled pressurizing. Make
sure that the exhaust pipeline cannot become blocked.
If there is an exhaust isolation valve, make sure that you
cannot operate the equipment with the valve closed to
avoid a risk of bursting!
+ Operation with silencer at the outlet:
Use the silencer only in case of low gas throughput.
Operating the pump at a high inlet pressure or pump-
ing dusty gases for a long time may cause clogging of
the silencer. Check the silencer regularly for perme-
ability and replace if necessary, or install a hose nozzle
instead.
+ Keep the electrical power cord away from heated sur-
faces.
Provide a rm, level platform for the equipment. Check
that the system which you are going to evacuate is
mechanically stable. Check that all ttings are secure.
Ensure a stable position of the pump without any me-
chanical contact other than the pump feet.
Comply with maximum permissible pressures at in-
let and outlet and with maximum permissible pres-
sure differences between inlet and outlet. See sec-
tion „Technical data“, pg. 23”. Do not operate the
pump with overpressure at the inlet.
Avoid overpressure of more than 17.5 psi absolute (1.2
bar absolute) in the event that gas or inert gas is con-
nected to the pump, or to a venting valve.
Note: Flexible elements will shrink when evacuated.
Connect hoses gas tight at inlet and outlet of the pump.