The vacuum gauge was adjusted using factory standards, which are traceable through
regular calibration in an accredited laboratory (German Calibration service) to the na-
tional standard. Depending on the process and/or accuracy requirements, check the
adjustment and readjust if necessary. For readjustment, the device has to be adjusted
both at atmospheric pressure as well as under vacuum.
The adjustment mode can be activated only if the process control is inactive. Press
key STOP if necessary.
Readjustment
Adjustment at atmospheric pressure
Ventilate the controller and/or the vacuum system. Make sure
that the vacuum connection at the controller is at atmospheric
pressure.
Press key por p simultaneously with key MODE, the
controller then switches to the adjustment mode.
The warning triangle and ”CAL are displayed for approx. 2
seconds.
While ”CAL is displayed, press key START/STOP.
Use key por p to adjust the reading to the actual at-
mospheric pressure.
Press key START/STOP to confirm.
Note: To determine the actual atmospheric pressure, use an accurate barometer or get accurate reading
from the weather service, the next airport.......(take into account the difference in altitude between e. g.
airport and laboratory).
p
+
CAL
Adjustment under vacuum
Evacuate the controller to a pressure < 0.5 mbar (e. g. by apply-
ing a good rotary vane pump).
Press key por p simultaneously with key MODE, the
controller then switches to the adjustment mode.
The warning triangle and ”CAL are displayed for approx. 2
seconds.
While ”CAL is displayed, press key START/STOP.
The reading is automatically adjusted to ”zero”.
Press key START/STOP to confirm.
Note: Adjustment under vacuum with an actual pressure higher than 0.5 mbar reduces the accuracy of the
measurement. If the pressure is significantly higher than 0.5 mbar, adjustment to a reference pressure is
recommended.
p
MODE
+
CAL
Adjustment at a reference pressure
Instead of adjustment under vacuum to a pressure < 0.5 mbar,
adjustment to a reference pressure within the range of 0 ..... 20
mbar is possible.
Press key por p simultaneously with key MODE, the
controller then switches to the adjustment mode.
The warning triangle and ”CAL are displayed for approx. 2
seconds.
While ”CAL is displayed, press key START/STOP.
The reading is automatically adjusted to ”zero”.
Use keys por p to adjust the display to the reference
pressure at the vacuum line within the range of 0 ..... 20
mbar.
Press key START/STOP to confirm.
Note: The accuracy of the value of the reference pressure will directly affect the accuracy of the adjustment.
If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum, the accuracy of the
controller might be doubtful. The diaphragm pump may not achieve the specified value (due to condensate,
poor state, failure of valves or the diaphragm).
p
MODE
+
CAL
p
0 .... 20 mbar
MODE