In case of low boiling solvents when the formation of condensate is unlikely, the use
of gas ballast might be unnecessary.
Shutdown
Short-term:
Has the pump been exposed to condensate?
- Allow the pump to continue to run at atmospheric pressure for a few minutes.
Has the pump been exposed to media which may damage the pump materials or
forms deposits?
- Check and clean pump heads if necessary.
Long-term:
- Take measures as described in section short-term shutdown.
- Separate pump from the application.
- Close gas ballast.
- Drain catchpots (if using with upgrade kits)
- Close inlet and outlet port (e. g. with transport caps).
- Store the pump in dry conditions.
In case of condensable vapours (water vapour, solvents,
...):
Do not pump vapour until the pump has reached its opera-
ting temperature.
Open gas ballast valve. The gas ballast valve is open if the
arrow on the gas ballast cap shows towards the pump.
With gas ballast valve open ultimate vacuum will be re-
duced, pumping speed is decreased.
Use inert gas at the air inlet to avoid the formation of explo-
sive mixtures.
Close gas ballast valve by turning 180°.
gas ballast
NOTICE