Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999198 / 29/09/2009
Use an open-ended wrench (w/f 17) to loosen the union
nut at the hose connection next to the gas ballast.
Use an open ended wrench (w/f 14) to turn the elbow t-
ting 1/4 of a turn, disconnect the hose. Do not remove the
elbow tting from the pump head.
Through reassembly a leak may result.
Use hex key size 5 to unscrew the handle from one side of
the housing cover.
Disassembling the housing cover at the inlet side
Lay the pumping unit down with the housing cover at the
side of the pump inlet facing upwards (support pumping
unit appropriately).
Use a hex key size 4 to loosen the six socket head screws
at the pump head and remove the upper housing (housing
cover with housing cover insert). Do not kink hoses.
Never remove parts using a spiky or sharp-edged tool (e.
g. screw driver), we recommend to use a rubber mallet or
compressed air (to be blown carefully into port).
Take the head covers carefully off the housing cover to
check the valves.
Note the position of the valves and remove them.
Replace valves if damaged.
Use petroleum ether or industrial solvent to remove de-
posits. Do not inhale.