User Manual

page 23 of 27
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999198 / 29/09/2009
If the pump does not achieve the ultimate vacuum:
In case the diaphragms and valves have been replaced, a run-in period of several hours is required
before the pump achieves its ultimate vacuum.
In case of loud noise switch off immediately and check diaphragms and valves for correct position.
If the specied ultimate vacuum is not achieved and if this does not change after the run-in period:
Check hose connectors between pump heads and manifolds for leaks. If necessary recheck pump
chamber.
Position housing cover. Pay attention to correct hose guid-
ance.
Move housing cover slightly to make sure that the head
covers are correctly positioned.
Screw in the six socket head screws xing the housing
cover crosswise rst slightly, then tighten.
Do not tighten until head cover is in contact with housing,
max. torque 6 Nm.
Assembling the housing cover at the side of the on/off switch
Slip the PTFE tube onto the hose connection at the hous-
ing cover.
Close hose clip with at pliers.
Tighten union nut rst by hand and then tighten one full
turn using open ended wrench.
Assemble the handle.
Use an open ended wrench (w/f 14) to reconnect the hose
to the elbow tting.