Technical data
* Information: In a motor speed range between 0 and 200 rpm the pump is running automatically in a
cycled interval operation.
Type PC 3001 basic
Maximum pumping speed (ISO 21360) m
3
/h 1.7
Ultimate vacuum (absolute) without gas ballast mbar 2
Ultimate vacuum (absolute) with gas ballast mbar 4
Maximum permissible inlet pressure (absolute) bar 1.1
Maximum permissible outlet pressure (abso-
lute)
bar 1.1
Maximum permissible pressure (absolute) at
gas ballast valve
bar 1.2
Permissible ambient temperature storage / op-
eration
°C -10 to +60 / +10 to +40
Permissible relative atmospheric moisture dur-
ing operation (no condensation)
% 30 to 85
Rated power (electric) kW 0.16
No-load speed* min
-1
0 - 2400
Maximum permissible range of supply voltage
( ±10% ) Attention: Observe specications
of rating plate!
100-120 V~ 50-60 Hz
200-230 V~ 50-60 Hz
Maximum rated current at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
1.6
0.7
Device fuse
1x 7A
1x 0.5A self-resetting (VACUU•BUS)
Motor protection
temperature sensor on the pcb
(current limitation)
Degree of protection IEC 529 IP 20
Inlet hose nozzle DN 6/10 mm
Outlet hose nozzle DN 10 mm / silencer
Dimensions L x W x H approx. mm 251 x 256 x 400