emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 101 of 120
Cleaning and inspecting the pump heads
➨ Remove catchpots (10) at inlet and outlet
(see „Use and operation“, pg. 38).
Detach the coupling of the hose connection
(M) below the housing cover (A).
➨ Open the hose clip (N) with a slotted screw-
driver.
➨ Pull the tubing off the hose connector.
Opening the hose clip:
➨ Apply slotted screwdriver as shown and
turn.
N
M
A
PC 3001 VARIO
pro
/ PC 3001 VARIO
pro
IK
Disassemble the exhaust vapor condenser.
➨ Loosen the union nut at the inlet of the ex-
haust vapor condenser.
➨ Remove the 4 screws afxing the counter
holder (O) of the exhaust vapor condenser
with a Torx driver T10.
➨ Remove the exhaust vapor condenser. In
doing so pull the PTFE hose out of the inlet
of the exhaust vapor condenser.
O