For replacing the diaphragm refer to the section Replac-
ing the diaphragm
above.
Assembling the housing cover at the side of the ON/OFF
switch
Put on housing cover (A).
Pay attention to a correct guidance of the connection hose towards the
inlet.
+ Move housing cover slightly to ensure that the head covers are cor-
rectly positioned.
Screw in the Allen head screws at the head
cover in a diagonal pattern with a 4 mm
wide Allen key, loosely at rst, to align, then
tighten.
+ Maximum torque: 4.4 ft
.
lb
f
(6 Nm).
Assembling the ttings
Attach the handle at the housing cover (use
5 mm wide Allen key)
Slip the hose onto the hose connector by
turning the tting (L) with an open-ended
wrench w/f 14.