+ Do not overll the condenser. Lowest permissible cool-
ant temperature: -112°F (-80°C).
+ Maximum operating pressure: 16 psi (1.1 bar) absolute.
Ensure that no parts of your clothing, hair or ngers
can be caught or drawn in at the inlet of the pump.
Never insert ngers or drop any other object into the
inlet or outlet.
Pumping at high inlet pressure may lead to overpres-
sure at the gas ballast valve. Pumped gases or con-
densate might be expelled if the valve is open. If an in-
ert gas supply is connected to the gas ballast, ensure
that its inlet pipeline is not contaminated.
You must take suitable precautions to prevent any
dangerous situation from arising if the controller starts
the pump, switches a coolant valve or opens a venting
valve.
Attention: If the controller is set to Autostart, the pro-
cess will start immediately after a power failure with-
out pressing any further key. It is your responsibility to
ensure that automatic start-up of the system will not
lead to any dangerous condition. Provide appropriate
safety measures. Check prior to starting the process
whether the option ”Autostart” (menu: conguration) is
enabled.
Pay attention to the safety symbol ”hot surfaces” on
the equipment. Hot parts may cause burns if touched.
Adopt suitable measures to prevent any danger aris-
ing from hot surfaces or electric sparks. Ensure that
hot surfaces of the pump do not cause burns. Provide
a suitable contact guard if necessary.
Ensure that the coolant outlet pipe at the waste vapor
condenser is always free and that it cannot get blocked.