emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 40 of 120
Use a suitable valve to isolate the pump from the vacuum
application. This is to allow the pump to warm up before
pumping condensable vapors and to clean the pump after
use before it is switched off.
When assembling, ensure vacuum-tightness. After as-
sembly, check the whole system for leaks.
Secure hose connections at the pump appropriately, e.g.,
with hose clamps, to protect against accidental detach-
ment.
To reduce pump noise emanating from the pump exhaust
port, connect an exhaust hose or use a silencer (see „Ac-
cessories“, pg. 94).
The VACUUBRAND controller CVC 3000 can only be
operated with components compatible with the VACU-
UBRAND VACUU•BUS system, (see „Accessories“, pg.
94).
The vacuum controller CVC 3000 controls VACUUBRAND
diaphragm pumps NT VARIO and VARIO-B, pumping
units PC 30xx VARIO, and optional coolant and venting
valves. Connected components (e.g., venting valve, level
sensor, external pressure transducer 3000 series) are au-
tomatically identied and congurated. Identical compo-
nents must be congurated beforehand; information upon
request.
Do not use more than one controller within the same
VACUU•BUS system.
Several controllers in the same VACUU•BUS system
will interfere with each other and result in error mes-
sages of the connected components (pumps, valves).
CVC 3000
The CVC 3000 is equipped with an internal capacitive
pressure transducer with ceramic diaphragm. It measures
the actual pressure independently of the gas type, and
with reference to the vacuum, i.e., absolute.