emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 45 of 120
Use the equipment only as intended, that is, for conden-
sation of vapors at the pump outlet or inlet.
Suitable coolants: e.g., dry ice or ethanol-dry ice mixture
or water-ice mixture.
➨ Remove cap. Fill coolant into the coolant
chamber. Put cap back in place. Do not
clamp the cap.
+ Check coolant level at appropriate intervals.
➨ Check level of condensate at appropriate
intervals and drain condensate if necessary.
+ Wait until the condensate has liqueed, if
necessary.
+ Separate the inner part of the condenser
from the vacuum installation to drain the
coolant.
➨ Remove the cap on top of the condenser.
Turn the inner part of the condenser and lift
to pull it out. Comply with regulations when
disposing of chemicals.
+ When reassembling check for the correct
position of the seal between inner and outer
parts of the condenser.
Notes concerning the operation of the inlet condenser
IK
+ Attention: The inlet condenser has no overpressure
safety relief device.
• Note: Install the hoses of the cooling system in a way
to avoid the ow / dripping of condensed water onto the
pumping unit (especially cables and electronic parts),
or use thermally insulated hoses to prevent condensa-
tion.
NOTICE
cap
inner part of