Use the equipment only as intended, that is, for conden-
sation of vapors at the pump outlet or inlet.
Suitable coolants: e.g., dry ice or ethanol-dry ice mixture
or water-ice mixture.
Remove cap. Fill coolant into the coolant
chamber. Put cap back in place. Do not
clamp the cap.
+ Check coolant level at appropriate intervals.
Check level of condensate at appropriate
intervals and drain condensate if necessary.
+ Wait until the condensate has liqueed, if
necessary.
+ Separate the inner part of the condenser
from the vacuum installation to drain the
coolant.
Remove the cap on top of the condenser.
Turn the inner part of the condenser and lift
to pull it out. Comply with regulations when
disposing of chemicals.
+ When reassembling check for the correct
position of the seal between inner and outer
parts of the condenser.
Notes concerning the operation of the inlet condenser
IK
+ Attention: The inlet condenser has no overpressure
safety relief device.
Note: Install the hoses of the cooling system in a way
to avoid the ow / dripping of condensed water onto the
pumping unit (especially cables and electronic parts),
or use thermally insulated hoses to prevent condensa-
tion.
NOTICE
cap
inner part of
condenser